AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 747
Entry time: Wed Oct 3 01:48:21 2018
|
Author: |
OH, TD |
Subject: |
Wednesday 3 October |
|
|
09.42 DAQ continues OK - current file Sep18/R2_36
09.42 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
nnaida2 rate due to one hot channel
Merger OK - see attachment 5
TapeServer OK - see attachment 6
no errors or warnings shown by server terminal sessions
1.8.L spectra - see attachments 7 & 8
1.8.W spectra 20us FSR - see attachments 9 & 10
09.46 ASIC check load (file Sep18/R2_36)
Note for p+n junction strip FEE64s the 1.8.W spectra are only updating for nnaida2, 10 & 17 - these traces
look OK
Other FEE64s show a noise transient and do not update - ASIC check load did not fix - probably requires
DAQ STOP/GO
09.57 temperature +25.4 deg C, RH 39.2%, d.p. +10.6 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level c. 80%
13:17 FEE temperatures and system wide checks all ok
Alpha run R2 stopped
R2_0 -> R2_37
13:18 Thresholds set to 0xa from 0x64 for all FEEs and ASICs
13:22 R3_0 started
Thresholds at 0xa and pulser on at 25Hz setting ~ 2Hz
ASIC settings saved to EXPERIMENTS/AIDA/2018Oct03-13.26.58
Merger screenshot - attachment 11
Stats - attachment 12
Layouts 7 and 8 - attachments 13 & 14
13:31 R3_0 Stopped
13:34 R4_0 Started
ASIC thresholds all at 0x9
Merger screenshot - attachment 15
Stats - attachment 16
Layout 1 - attachment 17
13:38 R4_0 Stopped
13:40 R5_0 Started
ASIC thresholds all at 0x8
Merger screenshot - attachment 18
Stats - attachment 19
Layout 1 - attachment 20
13:45 R5_0 Stopped
13:49 R6_0 Started
ASIC slow comp at 0x7
Merger screenshot - attachment 21
Stats - attachment 22
Layout 1 - attachment 23
13:52 R6_1 Stopped
R7_0 Started
ASIC Slow comp at 0x6
Merger screenshot - attachment 24
Stats - attachment 25
Layout 1 - attachment 26
Tapeserver window screenshot - attachment 27
Top window showing 72% CPU usage - attachment 28
15:06 Pulser measurements carried out with thresholds at 0xa
Typical pulser peak width was ~70-80 with a couple above 100.
Double peaking that was observed earlier in the week is no longer observed.
Both in the FEEs and on a scope
Backside pulser peaks - attachment 29
Frontside pulser peaks - attachment 30 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|