Sat Mar 29 11:14:51 2025
Author
*
:
Subject
*
:
> 08.40 DAQ continues OK - current file Oct18/R9_15 > > 08.42 all system wide checks OK *except* nnaida7 fails ADC calibration > > detector biases & leakage currents OK - see attachments 1 & 2 > > FEE64 temperatures OK - see attachment 3 > good event statistics OK - see attachment 4 > nnaida2 rate due to one hot channel > > Merger OK - see attachment 5 > TapeServer OK - see attachment 6 > no errors or warnings shown by server terminal sessions > > 1.8.L spectra - see attachments 7 & 8 > 1.8.W spectra 20us FSR - see attachments 9 & 10 > > 14.43 system wide checks ok, except nnaida7 failing adc calibration > > detector bias/leakage currents all ok (attachment 11) > > FEE64 temperatures ok (attachment 12) > > good event statistics ok > > no errors or warnings in terminal for Merger and TapeServer > > 15.05 Stopped alpha run - file R9_18 at 14.50 > lowered slow comparator threshold from 100 to 10 > good event statistics were ok, but with FEE1 at 400k (attachment 13) > did check on ASICs, stats were behaving ok (attachment 14) > took waveforms, were not initially triggering on p+n side. Change threshold from 8000 to 8500 (attachment 15, 16) > > 15.12 start new file (alpha run, R10) > changed threshold back to 100 from 10 > Merger and TapeServer ok > > > > >
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00