Entry time:
Sat Mar 29 12:36:49 2025
Author
*
:
Subject
*
:
09.58 DAQ continues OK - current file Oct18/R15_6 alpha background run 09.20 all system wide checks OK *except* nnaida7 fails ADC calibration SYNC errors since baseline - see attachment 1 - baseline not zero'd following reboot of nnaida2 and setup of all FEE64s yesterday SYNC errors OK following new baseline check detector biases & leakage currents OK - see attachments 2 & 3 FEE64 temperatures OK - see attachment 4 good event statistics OK - see attachment 5 nnaida2 rate due to one hot channel Merger OK - see attachment 6 TapeServer OK - see attachment 7 no errors or warnings shown by server terminal sessions 1.8.L spectra - see attachments 8 & 9 N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector ~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles 1.8.W spectra - see attachments 10 & 11 - trace updates for 9/12 p+n FEE64s have stopped, n+n FEE64s OK 15.54 DAQ continues - current file Oct18/R15_15 nnaida17 good event rate c. 400kHz 15.57 FEE64 reset, setup, system wide checks all ASIC clocks sync'd using ReSYNC all spectra zero'd ASIC check load DAQ go - data transfer disabled good events statistics - attachment 12 1.8.W spectra - see attachments 13 & 14 changed LED threshold 7700 -> 8500 for 9/12 p+n FEE64s