Entry time:
Sat Mar 29 20:13:19 2025
Author
*
:
Subject
*
:
09.37 DAQ continues OK - data transfer disabled FEE64 LEC/MEC slow comparator threshold 0xa 09.39 all system wide checks OK *except* nnaida7 fails ADC calibration detector biases & leakage currents OK - see attachments 1 & 2 FEE64 temperatures OK - see attachment 3 good event statistics OK *except* nnaida17 c. 400k - see attachment 4 nnaida2 rate due to one hot channel as before 1.8.L spectra - see attachments 5 & 6 N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector ~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles 1.8.W spectra 20us FSR - see attachments 7 & 8 10.10 LEC/MEC slow comparator from 0xa to 0x64 3x ASIC check load DAQ stop all histograms zero'd enable data transfer DAQ start - file Oct18/R16 beam background good event statistics OK *except* nnaida2 20kHz - see attachment 9