> 09.37 DAQ continues OK - data transfer disabled > FEE64 LEC/MEC slow comparator threshold 0xa > > 09.39 all system wide checks OK *except* nnaida7 fails ADC calibration > > detector biases & leakage currents OK - see attachments 1 & 2 > > FEE64 temperatures OK - see attachment 3 > good event statistics OK *except* nnaida17 c. 400k - see attachment 4 > nnaida2 rate due to one hot channel as before > > 1.8.L spectra - see attachments 5 & 6 > > N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector > ~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles > > 1.8.W spectra 20us FSR - see attachments 7 & 8 > > > 10.10 LEC/MEC slow comparator from 0xa to 0x64 > 3x ASIC check load > DAQ stop > all histograms zero'd > enable data transfer > DAQ start - file Oct18/R16 > > beam background > > good event statistics OK *except* nnaida2 20kHz - see attachment 9