> 06.53 DAQ continues OK - data transfer disabled > > 06.55 all system wide checks OK *except* nnaida7 fails ADC calibration > detector biases & leakage currents OK - see attachments 1 & 2 > Julabo FL11006 chiller set point +20 deg C actual +20 deg C, water level c. 75% (L-H) > ambient tempature + 26.9 deg C, RH 21.6%, d.p. +2.9 deg C > > FEE64 temperatures OK - see attachment 3 > good event statistics OK - see attachment 4 > > per FEE64 rate spectra - see attachment 5 > ASIC check load > per FEE64 rate spectra - see attachment 6 > good event statistics OK - see attachment 7 > > 1.8.L spectra - see attachments 8 & 9 > nnaida1 pulser peak width 72 ch FWHM > > 1.8.W spectra 20us FSR - see attachments 10 & 14 > 1.8.W spectra 200us FSR - see attachments 11 & 15 > 1.8.W spectra 2ms FSR - see attachments 12 & 16 > 1.8.W spectra 20ms FSR - see attachments 13 & 17 > Note - LED threshold 9000 for *all* FEE64s > > 07.27 statistics zero'd > some 32bit counters (e.g. nnaida19 & nnaida20) had overflowed/wrapped around > > 18:19 Computer appears to be running slow though cannot see a reason why. > No programs using to much ram or cpu > FEE temp ok - attachment 18 > bias and leakage currents ok - attachment 19 > System wide checks ok > > 21:02 System wide checks ok > DAQ Continues ok - data transfer disabled. > Bias and leakage currents ok - attachment 20 > FEE Temperatures ok - attachment 21
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown