Sat Mar 29 12:37:06 2025
Author
*
:
Subject
*
:
09.55 all system wide checks OK *except* nnaida7 fails ADC calibration FEE64 temperatures OK - see attachment 1 good event statistics OK - see attachment 2 all FEE64s <100k, 11x < 20k 1.8.L spectra - see attachments 3 & 4 nnaida1 70ch FWHM, nnaida3 80 ch FWHM detector biases & leakage currents OK - see attachments 5 & 6 1.8.W spectra 20us FSR - see attachments 7 & 8 nnaida1 Options, Local Controls, Data Redaout, Discriminator Controls - see attachment 9-13 Lost Activity Monitor -see attachment 14 18.01 DAQ found stopped This was caused by a chnage to the AIDA external clock frequency - see https://elog.ph.ed.ac.uk/AIDA/798 Following change of AIDA external clock frequency to 50MHz DAQ RESET DAQ SETUP Sync all ASIC clocks using Resync all system wide checks OK *except* nnaida7 fails ADC calibration zero all histograms DAQ GO 18.31 FEE64 temperatures OK - see attachment 15 good event statistics OK - see attachment 16 all FEE64s <100k, 10x < 20k detector biases & leakage currents OK - see attachments 17 & 18