Sat Mar 29 12:31:45 2025
Author
*
:
Subject
*
:
> > 09.55 all system wide checks OK *except* nnaida7 fails ADC calibration > > FEE64 temperatures OK - see attachment 1 > > good event statistics OK - see attachment 2 > all FEE64s <100k, 11x < 20k > > 1.8.L spectra - see attachments 3 & 4 > nnaida1 70ch FWHM, nnaida3 80 ch FWHM > > detector biases & leakage currents OK - see attachments 5 & 6 > > 1.8.W spectra 20us FSR - see attachments 7 & 8 > > nnaida1 Options, Local Controls, Data Redaout, Discriminator Controls - see attachment 9-13 > > Lost Activity Monitor -see attachment 14 > > 18.01 DAQ found stopped > > This was caused by a chnage to the AIDA external clock frequency - see https://elog.ph.ed.ac.uk/AIDA/798 > > Following change of AIDA external clock frequency to 50MHz > > DAQ RESET > DAQ SETUP > > Sync all ASIC clocks using Resync > > all system wide checks OK *except* nnaida7 fails ADC calibration > > zero all histograms > > DAQ GO > > 18.31 FEE64 temperatures OK - see attachment 15 > > good event statistics OK - see attachment 16 > all FEE64s <100k, 10x < 20k > > detector biases & leakage currents OK - see attachments 17 & 18