OH setup AIDA external clock using Tektronix AFG3102 - settings as follows CH1 ON - amplitude +5V, phase 0n deg, offset 0V, frequency 50.000000MHz JA setup DTAS/BigRIPS clock using Tektronix AFG3102 - settings as follows CH2 ON - amplitude -0.8V, phase 0n deg, offset 0V, frequency 25.000000MHz Unfortunately the AGF3102 was set to CH1 frequency = CH2 frequency and this changed the AIDA external clock frequency from 50MHz to 25MHz. This stopped the DAQ. Interestingly it was possible to Resync successfully and *all* system wide checks were OK *except* Check Clock Status which produced errors - see attachment 1. CH1 & CH2 now set to 50MHz and 25MHz respectively. AIDA DAQ now running OK.