Entry time:
Sat Mar 29 11:14:37 2025
Author
*
:
Subject
*
:
> 10.20 System wide checks ok (except naida07 ADC calibration failure) > > FEE temperatures ok (attachment 1) > > good event statistics ok (attachment 2) > > detector bias and leakage currents ok (attachment 3) > > DAQ stopped > > 10.40 pulser switched off > > slow comparator threshold raised to 0x64 > > DAQ, merger, TapeServer started - all ok (attachments 4 and 5) > > writing to file R1 in TapeData/May19 at 70 kb/s > > good event statistics (attachment 6) > > 16:11 Tapeserver and merger still ok > DAQ still running > System wide checks all ok *except nnaida7 fails ADC calibration* > Bias and leakage currents ok - attachment 7 > FEE Temperatures ok - attachment 8 > Statistics still ok > > Spectrum are showing alpha events (Layout 4) - attachment 9 > > 18:25 Tektronix clock accidentally unplugged. DAQ has stalled as a result. > Clock reset to 50MHz 5V TTL signal > > 18:29 System wide check of clock status shows error message in attachment 10 > > DAQ Reset > With slow comp at oxa statistics are still good - attachment 11 > System wide checks all ok *except nnaida7 fails ADC calibration* > Resync to the ASIC clock > All histograms zeroed > > 18:37 Slow comp threshold raised to 0x64 for alpha run > Writing to file: > May19/R2_0 > > > > > > > > >