are required
Entry time:
Sat Mar 29 12:25:53 2025
Author
*
:
Subject
*
:
> 12.47 DAQ continues OK to file May19/R7_3 > > alpha background > > ASIC settings 2018Oct03-13.26.58 > *except* slow comparator 0xa -> 0x64 > > BNC PB-4 Pulser OFF > > all system wide checks OK *except* nnaida7 fails ADC calibration > > detector biases & leakage currents OK - see attachments 1 & 2 > > FEE64 temperatures OK - see attachment 3 > good event statistics OK - see attachments 4 > > Merger OK - see attachment 5 > TapeServer OK - see attachment 6 > > 12.51 Check ASIC Control *all* ASICs & FEE64s > > 1.8.L spectra - see attachments 7 & 8 > p+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 9-12 > n+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 13-16 > > 13.13 analysis of May19/R7_3 - see attachment 17 > deadtime OK > ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104) > Also observed at a much lower rate for nnaida1 & nnaida2 > > 13.27 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up to H level marker > Temperature +24.6 deg C dew point +7.1 deg C RH 32.6%
Encoding
: