AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 805
Entry time: Sun May 19 02:21:43 2019
|
Author: |
CA |
Subject: |
Tests of low energy calibration on Alpha Data |
|
|
10.23 Low energy calibration program now produces relative gain coefficients in a format that can be used with AIDASort
using sorted alpha data Oct18/R29*
30 2.0GB files of data
LHS: Data sorted without relative gain-matching
RHS: Data sorted with relative gain-matching
Attachments (1-4) - DSSD1 - Ey-Ex
- Ey vs Ex
- Energy difference vs x strip
- Energy difference vs y strip
Attachments (5-8) - DSSD2 - Ey-Ex
- Ey vs Ex
- Energy difference vs x strip
- Energy difference vs y strip
Attachments (9-12) - DSSD3 - Ey-Ex
- Ey vs Ex
- Energy difference vs x strip
- Energy difference vs y strip
Attachments (13-16) - DSSD4 - Ey-Ex
- Ey vs Ex
- Energy difference vs x strip
- Energy difference vs y strip
Attachments (17-20) - DSSD5 - Ey-Ex
- Ey vs Ex
- Energy difference vs x strip
- Energy difference vs y strip
Attachments (21-24) - DSSD6 - Ey-Ex
- Ey vs Ex
- Energy difference vs x strip
- Energy difference vs y strip
Attachment 25 - Table comparing peak centroids and peak width (deviation, sigma) of Ey - Ex plots before and after gainmatching
- numbers found with simple gaussian fit
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Attachment 9: |
z2Ediff
57 kB
Uploaded Thu May 30 14:33:26 2019
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|