AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 81     Entry time: Wed Apr 29 11:34:09 2015
Author: Patrick Coleman-Smith 
Subject: Fast Trigger timings against Pulser Input and LEC/MEC Fast Comparator threshold setting 

I have tried the latest version ( 26 ) of the FEE64 code and carried out some measurements of the Fast Trigger delay from the Pulser signal ( not the Pulser Trigger out ) and observed the following.

The pulser gives a centroid of 28217 in the histogram.

        
Threshold setting Delay observed at MACB Fast Trigger output (nS)
0x40 no Trigger
0x38 no Trigger
0x30 no Trigger
0x28 1200 to 1800
0x20 690 to 860
0x18 505 to 600
0x10 320 to 385
0x8 210 to 270

This seems to me to be sensible as the discriminator is only a very simple comparator.

The OR64 ( 0xE )  signal works as True Fast NIM Logic. The OR16s ( 0xC ) logic works as inverse Fast NIM.

The OR64 is derived from the individual masked discriminator outputs. By default the Mask is not set.

 