AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 814
Entry time: Mon May 27 02:50:20 2019
|
Author: |
TD |
Subject: |
Monday 27 May |
|
|
10.50 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 60% M-H markers - topped up 1x jugful to 100% M-H markers
Temperature +25.6 deg C dew point +9.6 deg C RH 36.2%
DAQ continues OK to file May19/R11_8
alpha background
ASIC settings 2019May19-17.54.03
slow comparator 0x64
BNC PB-4 Pulser OFF
11.00 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
1.8.L spectra - see attachments 7 & 8
p+n FEE64s 1.8.W spectra 20us FSR - attachment 9
n+n FEE64s 1.8.W spectra 20us FSR - attachment 10
12.26 analysis of May19/R11_8 - see attachments 11
deadtime OK
11.05 DAQ stop file May19/R11_8
slow comparator 0x64 -> 0xa
Check ASIC Control *all* ASICs & FEE64s
good event statistics OK - see attachments 12
11x < 20k, 20x <50k, 23x <100k
11.25 DAQ start file May19/R12
Pulser walkthrough
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4
amplitude 90,000 - 10,000 @ 10,000 step
x5 attenuator IN
t_r 100ns tau_d 50us
frequency 250Hz range
delay MIN
tail pulse
- polarity
1.8.L spectra - see attachments 13 & 14
11.35 DAQ stop file May19/R12_1
11.48 DAQ start file May19/R13
Pulser walkthrough
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4
amplitude 12,000 - 2,000 @ 2,000 step
x5 attenuator IN
t_r 100ns tau_d 50us
frequency 250Hz range
delay MIN
tail pulse
- polarity
1.8.L spectra - see attachments 15 & 16
11.55 DAQ stop file May19/R13_1
16.20 AIDA moved into DTAS
Upstream edge of AIDA snout 17.7cm downstream of the upstream edge of the upstream DTAS NaI(Tl) detectors
which positions DSSSDs #3 & #4 at centre of DTAS. We will need to adjust the position of AIDA when we have
implant and PID-gated implant-decay correlation data to identify where specific isotopes stop.
17.05 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 80% M-H markers
Temperature +25.5 deg C dew point +9.3 deg C RH 35.9%
detector biases & leakage currents OK - see attachments 17 & 18
FEE64 temperatures OK - see attachment 19
good event statistics OK - see attachments 20
14x <20k, 22x <50k, 1x >100k
BNC PB-4
amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
frequency 25Hz range
delay MIN
tail pulse
- polarity
1.8.L spectra - see attachments 21 & 22
FEE FWHM
# (ch)
1 73
2 120
3 77
4 77
5 73
6 81
7 77
8 72
9 85
10 90
11 96
12 96
13 97
14 69
15 85
16 86
17 73
18 84
19 130
20 114
21 82
22 73
23 99
24 105
p+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 23-26
n+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 27-30 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|