Entry time:
Sat Mar 29 11:35:46 2025
Author
*
:
Subject
*
:
09.30 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C Water level c. 75% M-H markers Temperature +25.1 deg C dew point +9.7 deg C RH 37.9% DAQ continues OK all system wide checks OK *except* nnaida7 fails ADC calibration detector biases & leakage currents OK - see attachments 1 & 2 FEE64 temperatures OK - see attachment 3 good event statistics OK - see attachments 4 13x < 20k, 22x <50k, 24x <100k 1.8.L spectra - see attachments 5 & 6 nnaida1 73 ch FWHM, nnaida3 78 ch FWHM p+n FEE64s 1.8.W spectra 20us FSR - attachment 7 n+n FEE64s 1.8.W spectra 20us FSR - attachment 8