Entry time:
Sat Mar 29 12:26:55 2025
Author
*
:
Subject
*
:
> 14.10 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C > Water level c. 75% M-H markers > Temperature +25.7 deg C dew point +8.2 deg C RH 32.4% > > DAQ continues OK > > ASIC settings 2019May19-17.54.03 > slow comparator 0xa > > BNC PB-4 Pulser > Amplitude 90,000 > x5 attenuator IN > Frequency 25Hz range > t_r 100ns, tau_d 50us > - polarity > Delay MIN, tail pulse > > 14.14 all system wide checks OK *except* nnaida7 fails ADC calibration > > detector biases & leakage currents OK - see attachments 1 & 2 > > FEE64 temperatures OK - see attachment 3 > good event statistics OK - see attachments 4 > 11x <20k, 22x <50k, 24x <100k - 13x <20k yesterday > > 1.8.L spectra - see attachments 5 & 6 > nnaida1 73 ch FWHM, nnaida3 78 ch FWHM - no change cf. yesterday > p+n FEE64s 1.8.W spectra 20us FSR - attachment 7 > n+n FEE64s 1.8.W spectra 20us FSR - attachment 8 > FEE64 Rate spectra - attachment 9 > > Check ASIC Control *all* FEE64s & ASICs > zero all spectra > > 16.05 slow comparator 0x8 > good event statistics OK - see attachments 10 > 2x <20k, 11x <50k, 18x <100k > > slow comparator 0x9 > good event statistics OK - see attachments 11 > 6x <20k, 19x <50k, 23x <100k > > slow comparator 0xa > good event statistics OK - see attachments 12 > 10x <20k, 21x <50k, 24x <100k