AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 821
Entry time: Thu May 30 03:13:04 2019
|
Author: |
TD |
Subject: |
Thursday 30 May |
|
|
11.13 DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
10x < 20k, 21x < 50k, 24x < 100k - cf. yesterday 11x < 20k, 21x <50k
1.8.L spectra - see attachments 5 & 6
nnaida1 74 ch FWHM, nnaida3 78 ch FWHM - cf. yeserday nnaida1 73 -> 74 ch FWHM
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
1.8.L n+n spectra (log scale) no indication of light ions - attachment 9
N.B. currently no ions through ZDS and no increase in background for DTAS
This may/will change as online experiment starts to use ZDS from this evening
Check ASIC Control *all* FEE64s & ASICs
zero all spectra
FEE64 Rate spectra - attachment 10
Check ASIC Control
FEE64 ASIC
1 4
8 1
13 3
14 1
22 3
FEE64 Rate spectra - attachment 11
14.08 detector biases & leakage currents OK - see attachments 12 & 13
16.01 DAQ start (to TapeServer, no storage)
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
ADC data synchronised - see attachment 14
17.15 detector biases & leakage currents OK - see attachments 15 & 16 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|