11.31 DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
12x < 20k, 21x < 50k, 24x < 100k - cf. yesterday 10x < 20k, 21x < 50k
1.8.L spectra - see attachments 5 & 6
nnaida1 74 ch FWHM, nnaida3 78 ch FWHM - same as yesterday
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
no indication of light ions with beam to ZDS & F11
FEE64 Rate spectra - attachment 9
Check ASIC Control *all* FEE64s & ASICs
zero all spectra
17.21 detector biases & leakage currents OK - see attachments 10 & 11
FEE64 temperatures OK - see attachment 12
good event statistics OK - see attachments 13
11x < 20k, 19x < 50k, 21x < 100k - cf. 11.30 today 12x < 20k, 21x < 50k, 24x <100k
increased rates observed for nnaida17, 20 & 23 - all DSSSD#1 - nnaida22 ~15k -> ~30k
Check ASIC Control *all* FEE64s & ASICs
zero all spectra
cycled DSSSD#1 bias - no change
nnaida2 - no good events
unable to DAQ stop nnaida20 & nnaida23
unable to DAQ reset nnaida20 & nnaida23
obtain timeout errors from nnaida20 and DAQ error status for nnaida20 & nnaida23
reboot nnaida20 & nnaida23
DAQ reset
DAQ setup
ASIC settings 2019May19-17.54.03
slow comparator 0x64 -> 0xa
Check ASIC Control *all* FEE64s & ASICs
zero all spectra
still observe increased rates for nnaida17, 20 & 23 - all DSSSD#1 - nnaida22 ~32k
18.05 DAQ statistics return to ~normal - attachment 14
10x < 20k, 21x < 50k, 24x < 100k
11.31 17.21 18.05
nnaida17 13k 100k 43k
nnaida20 59k 136k 70k
nnaida22 16k 32k 26k
nnaida23 46k 111k 53k |