Message ID: 826
Entry time: Sun Jun 2 04:00:50 2019
|
Author: |
TD |
Subject: |
Sunday 2 June |
|
|
12.00 DAQ continues file May19/R15_169
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
10x <20k, 19x <50k, 24x <100k - same as yesterday evening
Merger OK - see attachment 5
TapeServer OK - see attachment 6
nnaida13 1.*.L spectra - attachment 7 - some evidence of light ions (5 < E < 20MeV) but low rate
FEE64 Rate spectra - attachment 8
1.8.L spectra - see attachments 9 & 10
nnaida1 75 ch FWHM, nnaida3 76 ch FWHM - cf. 77 & 82 ch FWHM respectively yesterday afternoon
p+n FEE64s 1.8.W spectra 20us FSR - attachment 11
n+n FEE64s 1.8.W spectra 20us FSR - attachment 12
Check ASIC Control *all* FEE64s & ASICs
12.20 Offline analysis from R15_0 to R15_167 ADC & disc data synchronisedfor file R15_0 - attachments 13, 14 & 15
18.27 DAQ stop file May19/R15_245
DAQ start - no storage
DAQ stall - reboot all FEE64s *except* nnaida4, 7, 8, 14, 15 & 16 - N.B. no power cycle required
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 19 & 20
FEE64 temperatures OK - see attachment 21
good event statistics OK - see attachments 16
8x <20k, 15x <50k, 21x <100k - cf. earlier today 10x <20k, 19x <50k, 24x <100k
p+n FEE64s 1.8.W spectra 20us FSR - attachment 17
n+n FEE64s 1.8.W spectra 20us FSR - attachment 18 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|