15.01 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C Water level c. 70% M-H markers - topup to 100% with 1x jugful water Temperature +25.8 deg C dew point +8.0 deg C RH 32.4% Access to F11 from 14.00 whilst F3 & F7 plastic scintillator and PPAC are replaced due to rad damage during online experiment DAQ continues OK ASIC settings 2019May19-17.54.03 slow comparator 0xa BNC PB-4 Pulser Amplitude 90,000 x5 attenuator IN Frequency 25Hz range t_r 100ns, tau_d 50us - polarity Delay MIN, tail pulse all system wide checks OK *except* nnaida7 fails ADC calibration detector biases & leakage currents OK - see attachments 1 & 2 FEE64 temperatures OK - see attachment 3 good event statistics OK - see attachments 4 13x <20k 21x < 50k 24x <100k cf. 15x <20k, 21x < 50k, 24x < 100k yesterday evening Merger OK - see attachment 5 TapeServer OK - see attachment 6 p+n FEE64s 1.8.W spectra 20us FSR - attachment 7 n+n FEE64s 1.8.W spectra 20us FSR - attachment 8 1.8.L spectra - see attachments 9 & 10 nnaida1 75 ch FWHM, nnaida3 82 ch FWHM - cf. 95 & 99 ch FWHM respectively yesterday afternoon FEE64 Rate spectra - attachment 11 Check ASIC Control *all* FEE64s & ASICs x2 FEE64 Rate spectra - attachment 12 - some channels recovered ADC data synchronised - attachment 11