Message ID: 829
Entry time: Wed Jun 5 06:57:38 2019
|
Author: |
TD |
Subject: |
Wednesday 5 June |
|
|
14.50 DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
13x <20k 21x < 50k 24x <100k cf. same yesterday afternoon
Merger OK - see attachment 5
TapeServer OK - see attachment 6
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
1.8.L spectra - see attachments 9 & 10
nnaida1 88 ch FWHM, nnaida3 92 ch FWHM - cf. 75 & 82 ch FWHM respectively yesterday afternoon
FEE64 Rate spectra - attachment 11
Check ASIC Control *all* FEE64s & ASICs x2
FEE64 Rate spectra - attachment 12 - some channels recovered
15.09 zero *all* spectra
zero DAQ stats
ADC data synchronised - attachment 11 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|