Message ID: 830
Entry time: Thu Jun 6 08:55:15 2019
|
Author: |
TD |
Subject: |
Thursday 6 June |
|
|
16.40 DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
13x <20k 21x < 50k 24x <100k cf. same yesterday afternoon
Merger OK - see attachment 5
TapeServer OK - see attachment 6
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
1.8.L spectra - see attachments 9 & 10
nnaida1 96 ch FWHM, nnaida3 103 ch FWHM - cf. 88 & 92 ch FWHM respectively yesterday afternoon - now
observe high amplitude tailing on pulser peaks
FEE64 Rate spectra - attachment 11
Check ASIC Control *all* FEE64s & ASICs x2
FEE64 Rate spectra - attachment 12 - some channels recovered
16.59 zero *all* spectra
zero DAQ stats
ADC data synchronised - attachment 11 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|