Message ID: 831
Entry time: Fri Jun 7 05:30:23 2019
|
Author: |
TD |
Subject: |
Friday 7 June |
|
|
14.05 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 70% M-H markers
Temperature +25.6 deg C dew point +10.3 deg C RH 37.7%
13.30 DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
13x <20k 21x < 50k 24x <100k cf. same yesterday afternoon
Merger OK - see attachment 5
TapeServer OK - see attachment 6
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
1.8.L spectra - see attachments 9 & 10
nnaida1 80 ch FWHM, nnaida3 87 ch FWHM - cf. 96 & 103 ch FWHM respectively yesterday aftenoon
FEE64 Rate spectra - attachment 11
Check ASIC Control *all* FEE64s & ASICs x2
FEE64 Rate spectra - attachment 12 - some channels recovered
14.20 BNC PB-4 pulser
Frequency changed from c. 4Hz to c. 2.5Hz (minimum)
25Hz range, variable fully anti-clockwise
14.33 zero *all* spectra
zero DAQ stats
ADC data synchronised - attachment 13
16.50 DAQ start file May19/R18
AIDA+DTAS merger test - 12kBq 60Co positioned on top of AIDA snout
DAQ sync issued by DTAS during R18_0
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
17.28 DAQ stop file May19/R18_4
17.40 DAQ start file May19/R19
AIDA+DTAS merger test - c. 30kBq? 137Cs positioned on top of AIDA snout
DAQ sync issued by DTAS during R18_0
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
18.18 DAQ stop file May19/R19_5
19.36 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 14 & 15
FEE64 temperatures OK - see attachment 16
good event statistics OK - see attachments 17
12x <20k 21x < 50k 24x <100k cf. 13x <20k 21x < 50k 24x <100k earlier this afternoon
19.45 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 70% M-H markers
Temperature +25.1 deg C dew point +7.9 deg C RH 33.6%
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|