Message ID: 833
Entry time: Sat Jun 8 06:08:52 2019
|
Author: |
CB, TD |
Subject: |
Saturday 8 June |
|
|
14.14 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 70% M-H markers
Temperature +25.3 deg C dew point +8.5 deg C RH 34.7%
DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
13x <20k 22x < 50k 24x <100k cf. 13x <20k 21x < 50k 24x <100k yesterday afternoon
Merger OK - see attachment 5
TapeServer OK - see attachment 6
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
1.8.L spectra - see attachments 9 & 10
nnaida1 104 ch FWHM, nnaida3 111 ch FWHM - cf. 80 & 87 ch FWHM respectively yesterday aftenoon
FEE64 Rate spectra - attachment 11
Check ASIC Control *all* FEE64s & ASICs x2
FEE64 Rate spectra - attachment 12 - some channels recovered
14.29 ADC data synchronised - attachment 13
14.31 zero all histograms
zero DAQ stats |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|