AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 842     Entry time: Sat Jun 15 01:05:41 2019
Author: CA, CB, TD, CG, OH, DK, ML 
Subject: Saturday 15 June 
09.05 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 75% M-H markers
      Temperature +24.1 deg C dew point +8.3 deg C RH 36.2%

      DAQ continues OK

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics - OK - see attachment 4
      11x <20k 23x < 50k 24x <100k cf. 9x <20k 23x < 50k 24x <100k 22.39 yesterday
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 74 ch FWHM, nnaida3 78 ch FWHM - cf. 81 & 77 ch FWHM respectively 23.20 yesterday

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      Check ASIC Control x2
      FEE64 ASIC
      10    4
      16    4
      17    4
      FEE64 Rate spectra - attachment 12 - some channels recovered

09.19 ADC data synchronised - attachment 13

09.23 zero all histograms
      zero DAQ stats

16:45 DAQ Continues to run
      Rates on 17, 20, 22 amd 23 are up - attachment 14
      Bias and leakage current ok - attachment 15
      System wide checks ok
      ASIC check ok
      
16:58 Statistics back to normal level - attachment 16

17.34 DAQ stop
      replaced kapton pcb on FEE no. 22

      DAQ start -> writing to May19/R36

      system wide checks ok, *except* naida7 fails adc calibration

      good event statistics ok - attachment 17

      TapeService and Merger ok - writing at 4 Mb/s - attachments 18 & 19

17.41 zeroed all histograms

c. 19.30 First beam to AIDA + DTAS - see attachments 20-24

      *no* degraders

      Observe most ions stop in DSSSD#2 with a few stopping in DSSSD#3
              focus looks OK but distribution is to LHS as you look upstream to BigRIPS

19.50 Removed upstream PPAC and adjusted ZDS transverse settings to centre - attachments 25-29

      Now observe ions now stop mostly in DSSSD#3 but significant fractions still stop ion DSSSD#2
  
      DSSSD#3 appears well focussed and centred

20.19 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachment 30

      FEE64 temperatures OK - see attachment 31

      good event statistics - OK - see attachment 32

20.53 xy hit patters in all DSSDs

      ions mostly in DSSD3, some still depositing in DSSD2

21.06 attachment 33 - beam slightly focused to the left - xy hit pattern

      attachment 34 - beam further focused - xy hit pattern

21.14 DAQ stop

      DAQ start writing to /RIBF147/R1

      Sync pulse sent by DTAS on file R1_0

21.17 DAQ stop/start

      

      writing to file /RIBF147/R2

      sync pulse sent by DTAS - file R2_0

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      no degrader

21.24 AIDA file on R2_1
      DTAS start 003
      BigRIPS run: 2003

21.31 histograms zeroed

      good event statistics ok - attachment 35

21.34 AIDA file on R2_2
      DTAS start 004
      BigRIPS run: 2004

22.02 good event statistics ok - attachment 36

22.16 AIDA file on R2_8
      DTAS run 004
      BigRIPS run 2004
      
      checked with DTAS
      
22.29 energy x vs y (HEC) - attachment 37 - grouping ~ 3MeV in all detectors
      xy hit pattern (HEC) - attachment 38
      implants centered on DSSD3

22.32 AIDA file R2_9
      DTAS run 005
      BigRIPS run 2005
      beam returns

22.34 HEC multiplicity - most events ~1,2,3 pixels - attachment 39

22.35 checked with DTAS
      AIDA file R2_10
      DTAS run 006
      BigRIPS run 2006

22.38 LEC energy x vs y - clear Ex=Ey locus - attachment 40
      xy hit pattern LEC - attachment 41
      xy multiplicity LEC - attachment 42 - most events 1,2 pixels

22.49 analyzer.f output for file R2_5 - attachment 44
      dead time 0.5% in fee19, zero in all others

22.56 problem with beam, beam stopped on AIDA file R2_8

23.05 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachment 45

      FEE64 temperatures OK - see attachment 46

      good event statistics - OK - see attachment 47

23.50 AIDA file R2_21
      DTAS run 006
      BigRIPS run 2006




      
      
      



      

 
      
Attachment 1: 1.png  12 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
1.png
Attachment 2: 2.png  12 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
2.png
Attachment 3: 3.png  135 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
3.png
Attachment 4: 4.png  85 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
4.png
Attachment 5: 5.png  304 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
5.png
Attachment 6: 6.png  98 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
6.png
Attachment 7: 7.png  159 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
7.png
Attachment 8: 8.png  146 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
8.png
Attachment 9: 9.png  119 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
9.png
Attachment 10: 10.png  121 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
10.png
Attachment 11: 11.png  151 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
11.png
Attachment 12: 12.png  150 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
12.png
Attachment 13: 13.png  56 kB  Uploaded Sat Jun 15 02:19:08 2019  | Hide | Hide all | Show all
13.png
Attachment 14: 190615_1646_stats.png  58 kB  Uploaded Sat Jun 15 09:46:55 2019  | Hide | Hide all | Show all
190615_1646_stats.png
Attachment 15: 190615_1647_bias.png  20 kB  Uploaded Sat Jun 15 09:47:48 2019  | Hide | Hide all | Show all
190615_1647_bias.png
Attachment 16: 190615_1658_stats.png  56 kB  Uploaded Sat Jun 15 09:59:10 2019  | Hide | Hide all | Show all
190615_1658_stats.png
Attachment 17: statsjun15.png  85 kB  Uploaded Sat Jun 15 10:39:43 2019  | Hide | Hide all | Show all
statsjun15.png
Attachment 18: tapejun15.png  98 kB  Uploaded Sat Jun 15 10:39:43 2019  | Hide | Hide all | Show all
tapejun15.png
Attachment 19: mergerjun15.png  97 kB  Uploaded Sat Jun 15 10:40:01 2019  | Hide | Hide all | Show all
mergerjun15.png
Attachment 20: 20.png  40 kB  Uploaded Sat Jun 15 12:45:32 2019  | Hide | Hide all | Show all
20.png
Attachment 21: 21.png  41 kB  Uploaded Sat Jun 15 12:45:32 2019  | Hide | Hide all | Show all
21.png
Attachment 22: 22.png  28 kB  Uploaded Sat Jun 15 12:45:32 2019  | Show | Hide all | Show all
Attachment 23: 23.png  115 kB  Uploaded Sat Jun 15 12:57:54 2019  | Hide | Hide all | Show all
23.png
Attachment 24: 24.png  118 kB  Uploaded Sat Jun 15 12:57:54 2019  | Hide | Hide all | Show all
24.png
Attachment 25: 30.png  31 kB  Uploaded Sat Jun 15 13:08:30 2019  | Hide | Hide all | Show all
30.png
Attachment 26: 31.png  42 kB  Uploaded Sat Jun 15 13:08:30 2019  | Hide | Hide all | Show all
31.png
Attachment 27: 32.png  43 kB  Uploaded Sat Jun 15 13:08:30 2019  | Hide | Hide all | Show all
32.png
Attachment 28: 33.png  117 kB  Uploaded Sat Jun 15 13:08:30 2019  | Hide | Hide all | Show all
33.png
Attachment 29: 34.png  124 kB  Uploaded Sat Jun 15 13:08:30 2019  | Hide | Hide all | Show all
34.png
Attachment 30: 190615_2018_bias.png  19 kB  Uploaded Sat Jun 15 13:20:23 2019  | Hide | Hide all | Show all
190615_2018_bias.png
Attachment 31: 190615_2014_temps.png  134 kB  Uploaded Sat Jun 15 13:20:36 2019  | Hide | Hide all | Show all
190615_2014_temps.png
Attachment 32: 190615_2018_stats.png  86 kB  Uploaded Sat Jun 15 13:20:44 2019  | Hide | Hide all | Show all
190615_2018_stats.png
Attachment 33: 100615_2051_xy.png  38 kB  Uploaded Sat Jun 15 13:52:51 2019  | Hide | Hide all | Show all
100615_2051_xy.png
Attachment 34: xy3.png  47 kB  Uploaded Sat Jun 15 14:05:44 2019  | Hide | Hide all | Show all
xy3.png
Attachment 35: xy4.png  44 kB  Uploaded Sat Jun 15 14:05:44 2019  | Hide | Hide all | Show all
xy4.png
Attachment 36: 2129_stats.png  85 kB  Uploaded Sat Jun 15 14:31:48 2019  | Hide | Hide all | Show all
2129_stats.png
Attachment 37: stats.png  86 kB  Uploaded Sat Jun 15 15:02:05 2019  | Hide | Hide all | Show all
stats.png
Attachment 38: epen.png  52 kB  Uploaded Sat Jun 15 15:24:16 2019  | Hide | Hide all | Show all
epen.png
Attachment 39: xy.png  54 kB  Uploaded Sat Jun 15 15:24:16 2019  | Hide | Hide all | Show all
xy.png
Attachment 40: multiHEC.png  28 kB  Uploaded Sat Jun 15 15:26:21 2019  | Hide | Hide all | Show all
multiHEC.png
Attachment 41: exeyLEC.png  44 kB  Uploaded Sat Jun 15 15:26:39 2019  | Hide | Hide all | Show all
exeyLEC.png
Attachment 42: xyLEC.png  98 kB  Uploaded Sat Jun 15 15:27:42 2019  | Hide | Hide all | Show all
xyLEC.png
Attachment 43: multiLEC.png  46 kB  Uploaded Sat Jun 15 15:28:34 2019  | Hide | Hide all | Show all
multiLEC.png
Attachment 44: R2_5  10 kB  Uploaded Sat Jun 15 15:49:07 2019  | Show | Hide all | Show all
Attachment 45: 2302_temps.png  134 kB  Uploaded Sat Jun 15 16:06:30 2019  | Hide | Hide all | Show all
2302_temps.png
Attachment 46: 2303_stats.png  86 kB  Uploaded Sat Jun 15 16:06:30 2019  | Hide | Hide all | Show all
2303_stats.png
Attachment 47: 2304_bias.png  19 kB  Uploaded Sat Jun 15 16:06:30 2019  | Hide | Hide all | Show all
2304_bias.png
ELOG V3.1.4-unknown