Entry  Thu Apr 30 15:28:25 2015, Patrick Coleman-Smith, High Energy Spectra from LYCCA with pulser LYCCA_High_Energy.png
    Reply  Thu Apr 30 17:43:36 2015, Patrick Coleman-Smith, High Energy Spectra from LYCCA with pulser 
       Reply  Thu Apr 30 20:21:20 2015, Patrick Coleman-Smith, High Energy Spectra from LYCCA with pulser 
Message ID: 86     Entry time: Thu Apr 30 15:28:25 2015     Reply to this: 88
Author: Patrick Coleman-Smith 
Subject: High Energy Spectra from LYCCA with pulser 

I have an eight FEE64 system now in place on the LYCCA chamber with a PB-5 pulser going into the adapter cards test input. The test capacitor is 30pF.

I was going to explore the effect of the thresholds on the pulser peak.

I notice that the spectra look like those taken in RIKEN and not a single peak. In the 1GeV range I would interpret this as noise but since this is such a high energy i don't quite understand what is happening in the ASIC.

I thought i would share it with you.

PB-5 input to 50R is 10v with no attenuation. Signal height on the inline 'scope is +5.2v

ASICs are set to "default" with 1GeV range and 3uS shaping time.

Let me know if anything useful can be gleaned by more tests on LYCCA.