Entry time:
Sat Mar 29 12:36:53 2025
Author
*
:
Subject
*
:
Database file - 2019Dec19-16.19.51 Current threshold at 0x64 13:27 Recovered from power cycle System wide checks all ok - Attachments 1-6 *NB* Synchronise the ASIC clocks doesn't work - Attachment 7 ASIC Control ok FEE temperatures ok - Attachment 8 Bias and leakage currents ok - Attachment 9 New taper service menu - Attachment 10 Pulser peaks log plot - Attachments 11-12 1.8.L pulser peaks - Attachment 13-14 Waveform plots - Attachment 15-16 FEE7 we aren't seeing a HIT pattern but we are seeing ADC data items -> It was a corrupted options file 15:20 DAQ Crashed after enabling data transfer Problem was not configuring the tapeserver correctly Have recovered from this. There are now no histograms for FEE09 This is because the options file has been corrupted during startup. From the statistics it appear we are still reading data on the FEE Chosen to let the DAQ run in this state to test stability for the time being