Author
*
:
Subject
*
:
> Database file - 2019Dec19-16.19.51 > Current threshold at 0x64 > > > 13:27 Recovered from power cycle > System wide checks all ok - Attachments 1-6 > *NB* Synchronise the ASIC clocks doesn't work - Attachment 7 > ASIC Control ok > FEE temperatures ok - Attachment 8 > > Bias and leakage currents ok - Attachment 9 > New taper service menu - Attachment 10 > > Pulser peaks log plot - Attachments 11-12 > 1.8.L pulser peaks - Attachment 13-14 > Waveform plots - Attachment 15-16 > > > > FEE7 we aren't seeing a HIT pattern but we are seeing ADC data items -> It was a corrupted options file > > 15:20 DAQ Crashed after enabling data transfer > Problem was not configuring the tapeserver correctly > Have recovered from this. There are now no histograms for FEE09 > This is because the options file has been corrupted during startup. > From the statistics it appear we are still reading data on the FEE > Chosen to let the DAQ run in this state to test stability for the time being