AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 865     Entry time: Wed Mar 1 16:42:06 2023
Author: Further results from Mezzanine bench tests 
Subject: PJCS 

Carried out some further work on failed mezzanine modules.

Inspected them externally and removed the copper block, internally with a bench microscope.

Found two obvious faults ..... broken components with scrambled pcb tracks.

To help determine problems with loading control registers I've added a new menu item in the ASIC control window to allow all four ASIC return values to be checked. This is only useful at the mezzanine test system.

Tested 7 that were labelled as faulty. Failed all of them. 2 with broken components and 1 with some broken bond wires. 

Tested one fresh from the cupboard. It failed with over current on the +5v psu ( > 4A! ).

I propose to create a method of testing these mezzanines which allows better access to the power supply regulators ( LT3080 ) and the associated control voltage.

 

Attachment 1: mezz_cap_gone_resized.jpg  330 kB  | Hide | Hide all
mezz_cap_gone_resized.jpg
Attachment 2: mezz_resistors_resized.jpg  363 kB  | Hide | Hide all
mezz_resistors_resized.jpg