AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
AIDA
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Sat Mar 29 11:18:15 2025
Author
*
:
Subject
*
:
> > Test of AIDA FEE64 d8:80:39:41:f6:ee ( aida01 ) - HDMI connector replaced. Cooling plate fitted - no base plate. > ISOL test station STFC DL, T9 > > Pulser BNC PB-5 > > Amplitude 0.5V > Attenuation x1 > Polarity + > Frequency 100Hz > Tail pulse, decay time 1ms > Delay 250ns (min) > > ASIC settings 2019May22-10.19.28 - attachment 1 > Slow comparator 0x80 > > Options - attachment 2 > > Adaptor PCB - PJCS > > > FEE64 temperatures OK - attachments 3 > > > System wide checks - attachments 4-9 > > FEE64 module aida01 global clocks failed, 6 > Clock status test result: Passed 3, Failed 1 > > FEE64 module aida01 failed > Calibration test result: Passed 3, Failed 1 > > Timestamp RAM values OK? > > > Statistics OK - attachments 10-11 > > Good WAVE events reported? > > > LEC spectra - attachments 12-15 > > Data observed for all channels of all ASICs - double peaking observed - may be due to incomplete FEE64 assembly? See attachment 17 > > > Hit/HitRate spectra - attachment 16 > > OK *except* > 1.2.L fast comparator missing > no WAVE data