AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 870
Entry time: Fri Jul 7 10:52:34 2023
|
Author: |
TD, PJCS |
Subject: |
Test of AIDA FEE64 d8:80:39:42:0d:0b ( aida03 ) - ASIC mezzanine replaced |
|
|
Test of AIDA FEE64 d8:80:39:42:0d:0b ( aida03 ) - ASIC mezzanine replaced
ISOL test station STFC DL, T9
Pulser BNC PB-5
Amplitude 1.0V
Attenuation x1
Polarity +
Frequency 100Hz
Tail pulse, decay time 1ms
Delay 250ns (min)
ASIC settings 2019May22-10.19.28
Slow comparator 0x80
Adaptor PCB - PJCS
FEE64 temperatures OK - attachments 1
System wide checks OK - attachments 2-5
Statistics OK - attachments 6-7
Hit/HitRate spectra OK - attachment 8
LEC spectra - attachments 9-12
Data observed for all channels of ASIC #1, 3, & 4
Data observed (double peaks) ASIC#2 channels 4-7 & 12-15, no data observed channels 0-3 & 8-11 - ASIC#2 address bit #2 set always ON? ASIC #2 fail or Mezzanine-FEE64
connection issue?
Typical pulser peak widths 15-17 channels FWHM ASIC #1-4
Wave spectra OK - attachments 13-16
19/7/23 removed mezzanine and cleaned contacts with ipa. cleaned FEE contacts. reassembled mezzanine to FEE and tested. ASIC 2 now ok. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|