Message ID: 880
Entry time: Wed Aug 2 08:44:00 2023
|
Author: |
TD |
Subject: |
Wednesday 2 August |
|
|
09.40 Al snout electrically connected ( < 1 Ohm ) to Al gasket and Al base plate
Al snout electrically isolated ( > 20 MOhm ) from aida02 FEE64 adaptor PCB and AIDA support frame
FEE64 temperatures OK - attachment 1
DSSSD bias +100V leakage current +7.565uA - attachment 2
All system wide checks OK
Slow comparator 0x64
per FEE64 rate spectra - attachment 3
per p+n FEE64 1.8.L spectra - attachments 4-5
aida01 pulser peak width 134 ch FWHM
per FEE64 1.8.W spectra 200us FSR - attachments 6-7
ADC data item stats - attachment 8
Slow comparator 0x14
ADC data item stats - attachment 9
per FEE64 rate spectra - attachment 10
Ground Al base plate to aida02 FEE64 adaptor PCB - appears to be marginally worse than without ground
per p+n FEE64 1.8.L spectra - attachments 11-12
aida01 pulser peak width 120 ch FWHM
per FEE64 1.8.W spectra 200us FSR - attachments 13-14
per FEE64 rate spectra - attachment 15
ADC data item stats - attachment 16
11.48 DSSSD bias +100V leakage current +11.055uA
Ground Al base plate to AIDA support frame - appears to be significantly worse than without ground
per FEE64 1.8.W spectra 200us FSR - attachments 17-18
per p+n FEE64 1.8.L spectra - attachments 19-20
aida01 pulser peak width 150 ch FWHM
per FEE64 rate spectra - attachment 21
ADC data item stats - attachment 22
11.48 DSSSD bias +130V leakage current +35.05uA
Ground Al base plate to aida02 FEE64 adaptor PCB - additional bias makes no significant difference
per FEE64 rate spectra - attachment 23
per p+n FEE64 1.8.L spectra - attachments 24-25
aida01 pulser peak width 118 ch FWHM
per FEE64 1.8.W spectra 200us FSR - attachments 26-27
ADC data item stats - attachment 28
14.03 DSSSD bias +100V leakage current +10.7uA
Ground Al base plate to aida02 FEE64 adaptor PCB - re-check
per FEE64 1.8.W spectra 200us FSR - attachments 28-29
ADC data item stats - attachment 30
15.10 Add DSSSD HV cable from aida04 to aida02 - significantly worse
Ground Al base plate to aida02 FEE64 adaptor PCB - re-check
per FEE64 1.8.W spectra 200us FSR - attachments 31-32
ADC data item stats - attachment 33
15.37 Switch from - to + polarity detector HV
N1419ET from ch #3 to ch #2, all outputs non-floating
LK1 ( local HV ground )removed from aida12, aida03, aida11 - installed aida04
LK3 ( DSSSD PCB ground ) installed aida03 ( no change )
HV #1 core aida12, aida03, aida11 - braid aida04
test + chain aida9, aida01, aida10, aida12, aida03, aida11 - connected to BNC PB-5 pulser
test - chain aida04, aida02 - not connected to pulser
AIDA PSU cabling per https://elog.ph.ed.ac.uk/AIDA/879
HV cable aida04 to aida02 - removed
Ground Al base plate to aida02 FEE64 adaptor PCB - removed
FEE64 temperatures OK - attachment 35
DSSSD bias -100V leakage current -7.155uA - attachment 36
Slow comparator 0x14
per FEE64 rate spectra - attachment 37
per p+n FEE64 1.8.L spectra - attachments 38-39
aida01 pulser peak width 123 ch FWHM but note double peaking
per FEE64 1.8.W spectra 200us FSR - attachments 40-41
ADC data item stats - attachment 42
Slow comparator 0x64
ADC data item stats - attachment 43
Very similar performance to + polarity DSSSD HV tests
16.05 Add LK1 to aida02
DSSSD bias -100V leakage current -7.770uA
Slow comparator 0x14
per FEE64 1.8.W spectra 200us FSR - attachments 44-46
ADC data item stats - attachment 47
per FEE64 rate spectra - attachment 48
Improved noise for n+n FEE64s esp. aida04 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|