Message ID: 882
Entry time: Thu Aug 3 09:21:26 2023
|
Author: |
TD |
Subject: |
Thursday 3 August |
|
|
10.20 DSSSD bias +100V leakage current +6.565uA - attachment 1
All system wide checks OK - attachments 2-7
why is SYNC rollover target 0xe?
FEE64 temps OK - attachment 8
per FEE64 1.8.W spectra 20us FSR - attachments 9-10
slow comparator 0x64
per FEE64 rate spectra - attachment 11
ADC data item stats - attachment 12
slow compararator 0x14
ADC data item stats - attachment 13
11.10 Test + & test - cables disconnected at the outputs of the BNC PB-5 and EG&G Ortec 534
slow comparator 0xf p+n rates dominated by 'hot' channels *except* aida03 & aida09
per FEE64 rate spectra - attachment 14
ADC data item stats - attachment 15
slow comparator 0xa p+n rates high ( 100-200k )
per FEE64 rate spectra - attachment 16
ADC data item stats - attachment 17
slow comparator 0x14 p+n rates entirely due to 'hot' channels
per FEE64 rate spectra - attachment 18
per FEE64 1.8.W spectra 20us FSR - attachments 19-20
ADC data item stats - attachment 21
11.27 Remove ground cable from Al base plate to aida02 FEE64 adaptor PCB
slow comparator 0xf p+n rates entirely due to 'hot' channels
per FEE64 1.8.W spectra 20us FSR - attachments 22-23
per FEE64 rate spectra - attachment 25
ADC data item stats - attachment 24
slow comparator 0xa p+n rates high ( 30-130k ) but lower than previously
per FEE64 rate spectra - attachment 26
ADC data item stats - attachment 27
11.48 DSSSD bias +100V leakage current +7.845uA
Remove LK3 ( local HV ground middle Si wafer )
slow comparator 0xa p+n rates high ( 55-135k ) but increase in rates appears to be due to 'hot' channels
per FEE64 1.8.W spectra 20us FSR - attachments 28-30
per FEE64 rate spectra - attachment 31
ADC data item stats - attachment 32
CAEN N1419ET ch #2 output connected to Tek TDS 210 DSO - powered from relay #6
ch 1 - ac - HV braid
ch 2 - ac - HV core
ch # 2 OFF - attachment 33
ch # 2 + 50V 0.000uA - attachments 34-35
with additional ThorLabs filter
ch # 2 + 50V 0.000uA - attachment 39
ch 2 - dc, hi Z
test - attachments 36-37
test + attachment 38
12.45 Install ThorLabs filter HV core ( *no* filter for HV braid )
slow comparator 0xa p+n rates high ( 40-130k ) but p+n rates dominated by 'hot' channels *except* aida03, aida11, aida12 where rates are generally higher per strip
per FEE64 rate spectra - attachment 40
per FEE64 1.8.W spectra 20us FSR - attachments 41-42
ADC data item stats - attachment 43
14.55 Install ThorLabs filter HV braid ( filters for both HV core & braid )
slow comparator 0xa p+n rates high ( 50-150k ) two filters probably somewhat worse than one filter
per FEE64 rate spectra - attachment 46
per FEE64 1.8.W spectra 20us FSR - attachments 44-45
ADC data item stats - attachment 47
Overall, effect of ThorLabs filters appears to belimited. Perhaps not unexpected as FEE64 adaptor PCBs also have a HV RC filter.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|