Entry time:
Sat Mar 29 12:37:01 2025
Author
*
:
Subject
*
:
17.00 USB-controlled ac mains relay - no water interlock ( yellow ) LED Water temperature ( station outside S4 entrance ) c. 19 deg C and pressure OK Flow can be felt on rubber outflow pipe from water manifold Failure of sensor or interlock box? *** Bypass interlock 16.58 Power ON 17.05 FEE64 temperatures OK - see attachment All system wide checks OK per FEE64 p+n 1.8.W spectra - attachments 2-42021Apr29-13-16-00 per FEE64 p+n 1,8.L spectra - attachment 5 aida01 pulser peak width 15 ch FWHM per FEE64 Rate spectra - attachment 6 BNC PB-5 pulser Amplitude 10.0V Attenuation x10 Decay 1ms Frequency 25Hz Polarity + ( n+n Ohmic test input from EG&G Ortec 534 Sum & Invert Amplifier not connected ) Clamp ON ASIC settings 2021Apr29-13-16-00 slow comparator 0x64 17.28 FEE64 temperatures - attachment 7 all < 1 deg C difference cf. 17.05 ADC data item stats OK - attachment 8 19.06 FEE64 temperatures OK - attachment 9