AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
AIDA
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Sat Mar 29 07:32:54 2025
Author
*
:
Subject
*
:
> > 15.06 Quick test of AIDA 24cm x 8cm 'triple' DSSSD before dismounting to install 1x bPlast detector downstream of the DSSSD > > Water flow, pressure and temperature OK > > Power Up OK > > All system wide checks OK > > DSSSD bias -150V leakage current -5.29uA > > ASIC settings 2021Apr29-13-16-00 > FEE64s re-configured to aida01 - aida08 by NH > aida02, aida04 negative input, all other FEE64s positive input > Currently unable to specify settings directory for Options so aida06 & auida08 are still defined as negative input FEE64s - need to change ASIC settings manually > > slow comparator p+n FEE64s 0x64->0xa, n+n FEE64s 0x64->0xf > > per FEE64 Rate spectra - attachment 1 > ADC data item stats OK - attachment 2 > p+n FEE64 rates dominated by 'hot' channels, other channels < 1Hz > n+n FEE64 rates ~ 10Hz > Known issues with aida04 asic #1 and aida10 ( now aida06 asic # 4 ) remain to addressed/investigated > > No chnages in performance cf. https://elog.ph.ed.ac.uk/AIDA/906 & https://elog.ph.ed.ac.uk/AIDA/907 > > >
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown