Message ID: 919
Entry time: Wed Nov 15 10:37:00 2023
|
Author: |
TD |
Subject: |
Wednesday 15 November |
|
|
11.36 Water flow, pressure & temperature ( 19 deg C ) OK
Power ON
Detector configuration - installed yesterday pm
Extant ( May 2022 campaign ) snout
1x MSL type BB18(DS)-1000 24cm x 8cm 'triple' DSSSD
1x ( downstream ) bPlast
2x additional Al foil sheets over end cap to ensure light tight ( zero dark current ) for bPlast
Dark current c 0.4uA measured for extant 2x Al foil sheets only
FEE64 configuration - aida01 - aida08
aida02 & aida04 n+n Ohmic strips, other FEE64s p+n junction strips
NH re-named FEE64s using dhcpd.conf - *no* changes to AIDA PSU cabling to FEE64s
All system wide checks OK
DSSSD bias -150V leakage current -5.140uA - attachment 1
FEE64 temps OK - attachment 2
ASIC settings 2021Apr29-13-16-00
Manually change ASIC settings aida06 & aida08 from - to + input
BNC PB-5 pulser to aida08
settings?
Slow comparator 0x64 all FEE64s
ADC data item stats - attachment 3
Increased rates and noise cf yesterday's test https://elog.ph.ed.ac.uk/AIDA/916
See attachments 4-9
Per FEE64 Rate spectra
Per FEE64 1.8.L spectra
Per FEE64 1.8.W spectra - 20us FSR
12.10 Disconnected power cables(?) from 3x bPlast cards around AIDA snout
1x card hanging to side of snout - aida04 side, 2x resting on top of AIDA snout ( inserted 2 sheets folded A4 paper between AIDA snout and PCBs )
Per FEE64 1.8.W spectra - 20us FSR - attachments 10 & 11
Per FEE64 Rate spectra - attachment 12
Significant improvement observed
Slow comparators 0xa p+n FEE64s, 0xf n+n FEE64s
pwer FEE64 1.8.W spectra - 20us FSR - attachments 13-15
ADC data item stats - attachment 16
per FEE64 Rate spectra - attachments 17 & 18
14.16 Removed bPlast cards from around snout - small improvment ADC stats
Attachments 19-22
15.10 aida01 FEE64 adaptor PCB re-seated ( c. 1 pin row out of position )
ASIC data item stats - attachment 23
Significant improvement which approaches previous noise performance
CC now assembling bPlast adaptor cards on a standoff Minitek frame
HA/IK have confirmed that bPlast cables are screened on one side only - CC demonstrated that surfaces of both sides of the bPlast cabling are non-conductive
Once assembled ( cards + power cables + input cables to bPlast detector ), observed increased noise similar to that observed this am
Attempt to return to status quo ante did not reduce noise but observed problems with aida01 ( asics #3 & #4 did not load ) - attachment 24
An unusual error message - power cycled and rebooted FEE64s
Slow comparator 0x64
ADC data item stats - attachment 27
Slow comparators 0xa p+n FEE64s, 0xf n+n FEE64s
ADC data item stats - attachments 26
per FEE64 Rate spectra - attachment 25
This appears to have ( largely ) restored status quo ante. Large number of hot channels but other typically close to previously observed rates.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|