Message ID: 936
Entry time: Sun Dec 10 13:06:27 2023
|
Author: |
TD |
Subject: |
Sunday 10 December |
|
|
14.05 S4 closed
Water flow and temperature ( gauges outside S4 entrance ) OK
BNC PB-5 pulser connected to p+n FEE64s by diasy chain of test + inputs
ASIC settings 2023Nov16-18.55.20
p+n FEE64s slow comparator 0xa, n+n FEE64s 0xf
All system wide checkks OK *except* adia08 ADC calibration failed
DSSSD bias -150V leakage current -5.28uA - OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
rates dominated by relatively few 'hot' channels
aida06 - no cables connected to FEE64 adaptor PCB
per FEE64 Rate spectra - attachment 4
aida06 - no cables connected to FEE64 adaptor PCB
system relatively quiet but has been quieter - no grounding of bPlast cabling and PSUs?
p+n and n+n FEE64 1.8.W spectra - 20us FSR - attachments 5-8
aida06 - no cables connected to FEE64 adaptor PCB
aida08 - no ADC calibration
18.25
SSSD bias -150V leakage current -5.100uA - OK - attachment 9
FEE64 temperatures OK - attachment 10
ADC data item stats - no significant change - attachment 11
rates dominated by relatively few 'hot' channels
aida06 - no cables connected to FEE64 adaptor PCB
per FEE64 Rate spectra - no significant change - attachment 12
aida06 - no cables connected to FEE64 adaptor PCB
system relatively quiet but has been quieter - no grounding of bPlast cabling and PSUs?
p+n and n+n FEE64 1.8.W spectra - 20us FSR - attachments 13-14
aida06 - no cables connected to FEE64 adaptor PCB
aida08 - no ADC calibration |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|