Author
*
:
Subject
*
:
> > At AIDA test station T9, STFC DL unable to start MIDAS DAQ with 1x FEE64 ( aida04 ) - see attachments > > aida04 appears to have booted OK and I can connect using telnet > > MIDAS 8015 server exits - signal 1 I’ve had this problem too. The solution is to have the single FEE twice in the list in the config file. The software is then quite happy. Alternatively a single FEE can still be accessed directly and will operate successfully. The only thing missing is the spectra function menu. They won’t work, so FWHM is harder to determine.