AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA, Page 1 of 46  ELOG logo
ID Date Author Subjectup Text Attachments
  75   Tue Apr 28 05:59:48 2015 AE"beta" trigger from MACB Trigger outputsWe've constructed a beta trigger with the
signals from the p-side of the three DSSDs
(OR of all fast
 6x 
  928   Wed Nov 22 15:12:40 2023 TD/tmp/log8015_error   log8015_error 
  282   Mon Jun 6 13:46:45 2016 TD, ML & YS/var/log/messagesaidas1 log  messages 
  14   Wed Oct 1 18:37:42 2014 Chris Griffin01/10/2014 Tests @ DLKept largely the same settings at the entry https://elog.ph.ed.ac.uk/AIDA/4
Same ASIC
settings
 11x 
  653   Sat Jun 10 15:56:31 2017 ZLiu0:00-8:00 11 June 2017
00:35   Stop DAQ
        AIDA file
RIBF128/R7_1172

00:36   Start DAQ
  
     AIDA file RIBF128/R7_1172
        BRIKEN
 46x 
  117   Wed Nov 11 17:36:19 2015 CG, TD11 Nov 2015 - new Kapton cable tests at DLUpon starting system with nothing connected
beyond adapter PCBs, observed pulser peak
FWHM of ~12-15ch in nnaida13+14 and ~15-18ch
 18x 
  118   Thu Nov 12 18:28:41 2015 CG, TD12 Nov 2015 - new Kapton cable tests at DLThis morning PC-S changed firmware and
DAQ software to that of the RIKEN setup
Firmware
 23x 
  119   Fri Nov 13 16:17:14 2015 CG, TD13 Nov 2015 - new Kapton cable tests at DLPowered system up with 4 bare 45cm kaptons
connected and nothing seems to have changed
from where we left last night
 6x 
  120   Sat Nov 14 20:19:43 2015 CG, TD, PJCS13 Nov 2015 - new Kapton cable tests at DL

    

       
            
  
  872   Sun Jul 16 09:36:40 2023 TD15-16 July 2023
Saturday 15 July
 40x 
  152   Tue Feb 16 17:32:21 2016 CG, TD16 February 2016MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.150uA, ambient temperature
+20.1 deg C
  
  153   Wed Feb 17 11:54:50 2016 CG, TD17 February 2016File R17 -> pulser walkthrough.
     - Started ~10.00
     - Pulser peak amplitude taken from 9.0V
 6x 
  649   Fri Jun 9 09:12:03 2017 ZLiu, AE17:00-24:0017:06 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R7_723
 leakage_jun9_6pm.pngstat_21.14.pngtemp2116.pngnnaida12_EL.pngnnaida9_EL.png 
  156   Fri Feb 19 10:23:31 2016 TD19 February 2016MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.070uA, ambient temperature
+20.2 deg C
  
  110   Fri May 22 11:58:02 2015 TD22 May 2015
MSL type BB18 serial 2998-22
 6x 
  228   Mon May 23 07:28:38 2016 TD23 May 201612.45 Rly16 'Sequence ALL On' operational

15.20 Startup remote control of CAEN N1419
 25x 
  231   Tue May 24 00:06:26 2016 TD24 May 2016Overnight. DAQ stopped. Detector bias off.
FEE power and recirculating chiller on.
 1.png2.png3.png4.png5.png 
  624   Fri Jun 2 08:25:34 2017 LHB2nd June Evening Shift16.34  Beam is off
       Aida R_1216, BRIKEN (83)/BigRIPS (1094)
continue to run for background data
 20x 
  233   Tue May 24 07:12:30 2016 TD3x WDC WD30EFRX ATA 3Tb disks added to aidas1 LVM configuration
The 3x WDC WD30EFRX ATA 3Tb disks installed
in aidas1 in April 2015 (formerly /data20,
 9x 
  630   Sun Jun 4 08:13:10 2017 LHB, PV, NN4th June 16.00-00.0016.25  *All* system wide checks OK *except*
ADC calibration of nnaida1, 21 and nnaida1
clock status (status=6)
 28x 
ELOG V3.1.4-unknown