Sun Jun 4 08:13:10 2017, LHB, PV, NN, 4th June 16.00-00.00 28x
|
16.25 *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
FEE64 Temperatures OK - see attachment 1
Good events statistics OK - see attachment 2
Detector biases & leakage currents OK - see attachment 3
No warnings/errors reported by merger since DAQ GO - see attachment 4
16.30 Analysis file RIBF148/R2_122 - nnaida2 deadtime 0.9%, all other deadtimes <0.3%, zero timewarps, one MBS
timewarp - see attachment 5
16.44 BRIKEN and BigRIPS stop daqs
AIDA on file R2_127
16.47 BRIKEN starts run 9
BigRIPS starts run 2008
AIDA on file R2_128
17.08 Online analysis - per DSSSD HEC x versus y attachment 6
per DSSSD HEC E_p versus E_n - see attachment 7
17.44 BRIKEN and BigRIPS stop daqs
AIDA on file R2_143
17.46 BRIKEN starts run 10
BigRIPS starts run 2009
AIDA on file R2_144
18.00 BRIKEN and BigRIPS stop daqs
AIDA on file R2_147
18.02 BRIKEN starts run 11
BigRIPS starts run 2010
AIDA on file R2_148
18.15 *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
FEE64 Temperatures OK - see attachment 8
Good events statistics OK - see attachment 9
Detector biases & leakage currents OK - see attachment 10
No warnings/errors reported by merger since DAQ GO - see attachment 11
18.47 Online analysis - per DSSSD HEC x versus y attachment 12
per DSSSD HEC E_p versus E_n - see attachment 13
Zero online sort spectra
18.50 Analysis of file RIBF148/R2_159 - all deadtimes <0.5%, zero timewarps, four MBS timewarps - see attachment 14
19.04 BRIKEN and BigRIPS stop daqs
AIDA on file R2_165
19.07 BRIKEN starts run 12
BigRIPS starts run 2011
AIDA on file R2_166
20:00 BRIKEN and BigRIPS stop daqs
AIDA on file R2_180
20.02 BRIKEN starts run 13
BigRIPS starts run 2012
AIDA on file R2_180
Zero online sort spectra so we can monitor the influence on HEC x versus y due to the degrader just inserted.
20.15 *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
FEE64 Temperatures OK - see attachment 15
Good events statistics OK - see attachment 16
Detector biases & leakage currents OK - see attachment 17
No warnings/errors reported by merger since DAQ GO - see attachment 18
20.28 Analysis of file RIBF148/R2_185 - all deadtimes <0.4%, zero timewarps, three MBS timewarps - see attachment 19
20.58 Online analysis - per DSSSD HEC x versus y attachment 20
per DSSSD HEC E_p versus E_n - see attachment 21
Zero online sort spectra
21.02 BRIKEN and BigRIPS stop daqs
AIDA on file R2_197
21.03 BRIKEN starts run 14
BigRIPS starts run 2013
AIDA on file R2_197
22.02 BRIKEN and BigRIPS stop daqs
AIDA on file R2_213
22.03 BRIKEN starts run 15
BigRIPS starts run 2014
AIDA on file R2_214
22.17 *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
FEE64 Temperatures OK - see attachment 22
Good events statistics OK - see attachment 23
Detector biases & leakage currents OK - see attachment 24
No warnings/errors reported by merger since DAQ GO - see attachment 25
22.22 Analysis of file RIBF148/R2_218 - all deadtimes <0.5%, zero timewarps, four MBS timewarps - see attachment 26
22.45 Online analysis - per DSSSD HEC x versus y attachment 27
per DSSSD HEC E_p versus E_n - see attachment 28
Zero online sort spectra
23.04 BRIKEN and BigRIPS stop daqs
AIDA on file R2_230
23.05 BRIKEN starts run 16
BigRIPS starts run 2015
AIDA on file R2_231 |
Tue May 24 07:12:30 2016, TD, 3x WDC WD30EFRX ATA 3Tb disks added to aidas1 LVM configuration 9x
|
The 3x WDC WD30EFRX ATA 3Tb disks installed in aidas1 in April 2015 (formerly /data20, /data30 and
/data40) have now been added to the aidas1 logical volume management (LVM) configuration to form a
single c. 8Tb filesystem (/data10) which will be automatically mounted at boot.
N.B. Although the disks are 3Tb capacity, Linux OS/BIOS can only access the first 2.2Tb of each disk.
Data previously stored on /data20, /data30 and /data40 has been copied to the local Seagate USB 8Tb
disk (/ntfs) and the University of Edinburgh ECDF multi-Pb disk array.
aidas1> df -h
Filesystem Size Used Avail Use% Mounted on
/dev/mapper/vg_aidas1-lv_root
50G 27G 20G 58% /
tmpfs 3.9G 1.1M 3.9G 1% /dev/shm
/dev/sdb1 477M 77M 376M 17% /boot
/dev/mapper/vg_aidas1-lv_home
860G 106M 816G 1% /home
/dev/mapper/Data1-data10
7.7T 1.6T 5.8T 22% /data10
/dev/sda2 7.3T 3.4T 4.0T 46% /ntfs
[root@aidas1 /]# ln -s /data10/TapeData /TapeData
[root@aidas1 /]# ls -l /
total 210
dr-xr-xr-x. 2 root root 4096 Apr 7 03:49 bin
dr-xr-xr-x. 5 root root 1024 Feb 17 2015 boot
drwxr-xr-x. 10 root root 4096 Jul 23 2015 cgroup
drwxr-xr-x. 8 root root 4096 Dec 12 19:29 data10
drwxr-xr-x. 2 root root 4096 Apr 28 2015 data20
drwxr-xr-x. 2 root root 4096 Apr 28 2015 data30
drwxr-xr-x. 2 root root 4096 Apr 28 2015 data40
drwxr-xr-x. 21 root root 4500 May 24 14:40 dev
drwxr-xr-x. 135 root root 12288 May 24 17:55 etc
drwxr-xr-x. 4 root root 4096 Apr 28 2015 home
drwxr-xr-x. 4 root root 4096 Jan 8 2014 homes
dr-xr-xr-x. 13 root root 4096 Feb 18 03:19 lib
dr-xr-xr-x. 11 root root 12288 Apr 14 06:02 lib64
drwx------. 2 root root 16384 Jan 7 2014 lost+found
drwxr-xr-x. 2 root root 4096 May 24 11:00 media
lrwxrwxrwx. 1 root root 12 Apr 22 2014 MIDAS -> /MIDAS@aidas
drwxrwxr-x. 29 npg npgstaff 4096 May 3 23:47 MIDAS@aidas
drwxr-xr-x. 22 root root 4096 Dec 12 12:10 MIDAS@aidas.121215
drwxr-xr-x. 28 root root 4096 Feb 24 2015 MIDAS@aidas.240215
drwxr-xr-x. 2 root root 0 May 24 11:00 misc
drwxr-xr-x. 2 root root 4096 Jul 21 2011 mnt
drwxr-xr-x. 2 root root 0 May 24 11:00 net
drwxrwxrwx. 1 root root 4096 Apr 28 10:55 ntfs
drwxr-xr-x. 4 root root 4096 Feb 27 2014 opt
dr-xr-xr-x. 411 root root 0 May 24 10:58 proc
dr-xr-x---. 28 root root 4096 May 24 18:01 root
drwxr-xr-x. 2 root root 12288 Feb 18 03:19 sbin
dr-xr-xr-x. 2 root root 4096 Feb 27 2014 scratch
drwxr-xr-x. 7 root root 0 May 24 10:58 selinux
drwxr-xr-x. 2 root root 4096 Jul 21 2011 srv
drwxr-xr-x. 13 root root 0 May 24 10:58 sys
lrwxrwxrwx. 1 root root 16 May 24 18:02 TapeData -> /data10/TapeData
drwxrwxrwt. 41 root root 65536 May 24 18:01 tmp
drwxr-xr-x. 13 root root 4096 Jan 7 2014 usr
drwxr-xr-x. 23 root root 4096 Jan 7 2014 var
dr-xr-xr-x. 2 root root 4096 Feb 27 2014 WWW |
Fri Jun 2 08:25:34 2017, LHB, 2nd June Evening Shift 20x
|
16.34 Beam is off
Aida R_1216, BRIKEN (83)/BigRIPS (1094) continue to run for background data
16.40 BRIKEN run 83 and BigRIPS run 1094 stopping daqs
AIDA on file R1_1218
16.44 Still no beam, running background
BigRIPS starts run 1095
BRIKEN starts run 84
Aida on file R1_1218
17.55 Still no beam, running background
System wide checks all ok
FEE temperatures all ok - attachment 1
Good events statistics - attachment 2
Bias and leakage currents ok - attachment 3
Merger ok - attachment 4
17.58 Beam on, AIDA file R1_1234
17.59 Briken run 84 and BigRIPS run 1095 stopping daqs
AIDA on file R1_1234
18.00 BigRIPS starts run 1096
BRIKEN starts run 85
Aida on file R1_1234
Zero'd online SAS spectra
19.00 BRIKEN run 85 and BigRIPS run 1096 stopping daqs
AIDA on file R1_1248
19.02 BigRIPS starts run 1097
BRIKEN starts run 86
Aida on file R1_1248
19.04 Online analysis: DSSSD HEC x versus y spectra - see attachment 5
20.00 BRIKEN run 86 and BigRIPS run 1097 stopping daqs
AIDA on file R1_1261
20.01 BigRIPS starts run 1098
BRIKEN starts run 87
Aida on file R1_1261
20.04 Beam is off
Aida R_1262, BRIKEN (87)/BigRIPS (1098) continue to run for background data
20.15 System wide checks all ok
FEE temperatures all ok - attachment 6
Good events statistics - attachment 7
Bias and leakage currents ok - attachment 8
Merger ok - attachment 9
20.21 BRIKEN run 87 and BigRIPS run 1098 stopping daqs because beam about to be switched back on
AIDA on file R1_1265
20.22 Beam on
BigRIPS starts run 1099
BRIKEN starts run 88
Aida on file R1_1266
21.18 Online analysis: DSSSD HEC x versus y spectra - see attachment 10
Zero'd SAS spectra
21.19 BRIKEN run 88 and BigRIPS run 1099 stopping daqs
AIDA on file R1_1278
21.20 BigRIPS starts run 1100
BRIKEN starts run 89
Aida on file R1_1279
22.05 System wide checks all ok
FEE temperatures all ok - attachment 11
Good events statistics - attachment 12
Bias and leakage currents ok - attachment 13
Merger ok - attachment 14
22.17 Analysis of file RIBF127R1/R1_1290 - dead time < 0.4% - see attachment 15
22.18 BRIKEN run 89 and BigRIPS run 1100 stopping daqs
AIDA on file R1_1291
22.20 BigRIPS starts run 1101
BRIKEN starts run 90
Aida on file R1_1292
22.24 Online analysis: DSSSD HEC x versus y spectra - see attachment 16
Zero'd SAS spectra
23.20 BRIKEN run 90 and BigRIPS run 1101 stopping daqs
AIDA on file R1_1305
23.22 BigRIPS starts run 1102
BRIKEN starts run 91
Aida on file R1_1305
23.48 System wide checks all ok
FEE temperatures all ok - attachment 17
Good events statistics - attachment 18
Bias and leakage currents ok - attachment 19
Merger ok - attachment 20 |
Tue May 24 00:06:26 2016, TD, 24 May 2016    
|
Overnight. DAQ stopped. Detector bias off. FEE power and recirculating chiller on.
08.00 Detector bias on OK - see attachments 1-2
DAQ start OK - see attachments 3-4
All system-wide tests OK *except* nnaida9 and nnaida29 failed ADC calibration
Temperatures OK
18.00 Saved corrected ASIC settings to DB Key 2016May14-13.57.38
Changed options - nnaida5 master timestamp sync phase from 0x1 to 0x3 |
Mon May 23 07:28:38 2016, TD, 23 May 2016 25x
|
12.45 Rly16 'Sequence ALL On' operational
15.20 Startup remote control of CAEN N1419 4-ch HV modules
DSSSD #1-7 leakage currents OK - see attachments 1-2
Ambient temperature +24.8 deg C
15.35 Plot Temperature appears to plot the first 15,000 (or so) temperature readings
and not the *most recent* 15,000 (or so) readings. See attachment 3.
There is no way to change the time window for the temperature plot
As a workaround, empty file using the command
cp /dev/null /MIDAS/log/temperatures.txt
17.10 For all FEEs (*except nnaida19 & nnaida26) which are currently disconnected)
reprogram FEE flash memory using FEE_GF_Feb16_17.bin
Power cycle
attempt #1 - wait 4'
STATE for nnaida9 returned with an error
connect failed
NONE
connect failed
attempt #2 - wait 6'
STATE for nnaida4 returned with an error
connect failed
NONE
connect failed
attempt #3 - wait 8' - OK
Installed updated Check software - Check.tml .tcl .js and Common.js
Started DAQ - histogramming enabled - data transfer disabled
Large fraction of FEEs not producing SYNCs, SYNC rate low (~20-30Hz) for other FEEs - see attachments 4 & 5
Note slow comparator 0xc8 (i.e. high) and fast discriminators (LEC/MEC) disabled
Firmware version - attachment 6
System wide checks ASIC clock sync (30/30 fail) and ADC calibration (7/30 fail) work as before but
new checks produce errors - see attachments 7-11
Additional screenshots from Master Timestamp Control - see attachments 12-17
00:40 PJCS identifies workaround to above problems - correlation controls now enabled for master timestamp
FEE (nnaida5) only. May also explain issues with nnaida19 & nnaida26.
PJCS will produce detailed, separate Elog entries
Currently, all (30) FEE64s running OK with sensible AIDA SYNC rates.
But. No spare fuses (16A/250V slow blow) for USB-controlled ac mains relay. Until Chris arrives in
c. 9h. We will explore options to use *second* USB-controlled ac mains relay. |
Fri May 22 11:58:02 2015, TD, 22 May 2015 6x
|
MSL type BB18 serial 2998-22
CAEN N1419B floating HV bias +150V I_L +4.255uA
(bias network locally grounded at AIDA adaptor PCB connected to p+n junction strips)
ASIC settings attachments 1-4
BNC PB-5 Pulser
Amplitude 0.50V
Attenuator x1
Polarity - (output split to Cooknell SA2 to provide + polarity signals)
Decay time 1ms
Frequency 50Hz
Unable to replace PCL Kapton PCBs with FCT Kapton PCBs.
FCT Kapton PCBs exhibited either a high resistance short
(15v/50uA) or expecte diode response as voltage increased
from zero until ~20-30V when leakage curent abruptly increased
to CAN N1419B OVC limit (curremtly 50uA). Return FCT Kapton PCBs
to Edinburgh for continuity/short tests in the automated tester
in the electronics workshop.
MSL type BB18 serial 2998-20
CAEN N1419B floating HV bias +200V I_L +5.705uA |
Fri Feb 19 10:23:31 2016, TD, 19 February 2016
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.070uA, ambient temperature +20.2 deg C
DSSSD - AIDA adaptor PCB cabling
4 off LH Coupler (Kapton PCB, 5cm), 2x 34-way Samtec ribbon cable (45cm), RH coupler(Kapton PCB, 10cm)
+ 3M 1245 1.4mil copper foil screen ribbon cables + RH coupler only (i.e. not LH coupler)
+ drain wires -> gold-plated Lemo-00 test input connectors
nnaida 11 & 12 AIDA adaptor PCB rev B
nnaida 13 & 14 AIDA adaptor PCB rev C (LK1 fitted)
ground links LK3 & LK7 fitted to nnaida11-14 AIDA adaptor PCBs
Heavy duty copper cable connects copper front end frames of FEE modules
Nitto 5011N conductive gasket between FEE module and front end frames
Standard ASIC settings
nnaida 11 & 12 - negative input
nnaida 13 & 14 - positive input
Pulser BNC PB-5
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 5.00000V
Polarity +
Pulse top Tail
Atten 10x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier
207Bi source, approx centred on DSSSD, approx 3cm from DSSSD
Compare measurements to previous electronic noise tests 27 January 2016
See https://elog.ph.ed.ac.uk/AIDA/145
nnaida11
Shaping time(us)
0.5 1 2 4 8
87 80 75 60 54
88 81 76 62 58
87 83 78 62 57
84 74 69 56 56
81 78 72 59 56
86 79 74 60 56
85 73 65 56 57
85 77 72 59 57
83 77 73 60 56
85 81 77 62 58
85 71 65 55 55
80 76 72 59 56
85 78 73 59 57
87 76 71 58 55
93 85 78 62 55
91 86 81 64 55
85.8 78.4 73.1 59.6 56.1 mean
3.2 4.2 4.5 2.5 1.2 std dev
nnaida12
0.5 1 2 4 8
87 77 69 58 55
83 78 75 61 58
80 84 77 63 56
86 73 64 54 54
85 77 69 58 55
86 77 71 59 55
85 72 64 55 55
84 74 66 56 55
82 74 67 57 55
80 78 73 61 55
80 70 63 55 55
82 72 65 56 53
82 73 69 58 56
90 74 64 55 55
95 84 74 59 53
86 82 74 59 55
84.6 76.2 69.0 57.8 55.0
4.0 4.3 4.6 2.5 1.2
nnaida13
0.5 1 2 4 8
51 51 51 46 47
49 53 52 47 48
53 60 61 50 49
50 46 45 42 46
48 50 50 45 48
57 54 54 47 49
49 47 46 44 48
49 50 50 46 49
50 51 52 46 48
50 54 54 47 49
49 45 44 43 48
47 50 50 45 47
49 52 52 47 49
48 45 42 42 47
54 55 55 47 45
47 54 53 47 46
50.0 51.1 50.7 45.7 47.7
2.7 4.0 4.7 2.1 1.3
nnaida14
0.5 1 2 4 8
81 79 75 60 53
79 77 73 57 52
79 78 73 61 54
85 81 76 61 54
79 71 70 56 51
82 79 74 60 54
83 78 73 59 54
79 74 70 56 52
82 78 73 59 54
76 75 70 54 52
82 75 72 58 54
81 76 71 58 54
78 75 70 58 53
96 87 81 62 54
108 103 94 69 55
95 90 85 68 54
84.1 79.8 75.0 59.8 53.4
8.5 7.8 6.5 4.0 1.1
* outlier ignored
** ground links LK3 & LK7 fitted to nnaida11-14 AIDA adaptor PCBs
electronic noise
p+n *and* n+n strips ~ 55 ch FWHM ~ 0.7 * 55 ~ 39keV FWHM
=> slow comparator (5x sigma) ~ 5 * 39 / 2.355 ~ 82keV
cf.
See https://elog.ph.ed.ac.uk/AIDA/145
n+n strips ~64 ch FWHM ~ 0.7 * 64 ~ 45keV FWHM
=> slow comparator threshold (5x sigma) ~ 5 * 45 / 2.355 ~ 96keV
p+n strips ~ 54 ch FWHM ~ 0.7 * 54 ~ 38keV FWHM
=> slow comparator threshold (5x sigma) ~ 5 * 38 / 2.355 ~ 81keV
Pulser BNC PB-5
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 10.00000V
Polarity +
Pulse top Tail
Atten 10x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier
207Bi source, approx centred on DSSSD, approx 3cm from DSSSD
File R35
shaping time 8uS
slow comparator 6 (dec)
start: 13.26 stop: 13.50
spectra saved to disk 14.52 BST (sic)
Pulser amplitude 10.0000V-1.00000V @ 1.00000V step |
Fri Jun 9 09:12:03 2017, ZLiu, AE, 17:00-24:00    
|
17:06 BRIKEN and BigRIPS stop their DAQs
AIDA on file R7_723
17:08 BigRIPS starts run 3100
BRIKEN starts run 94
AIDA on file R7_724
18:11 Stop BRIKEN/BigRIPS run. AIDA at R7_739
18:13 Start new run: BRIKEN 95, BigRIPS run 3103
AIDA run R7_740
Beam stopped right when the run was started.
18:17 Beam back, with intensity about 49 pnA
18:43 AIDA leakage currents [uA]: 4.665 5.325 5.705 5.125 5.05 4.25
19:09 BRIKEN starts run 096
BigRIPS starts run 3102
AIDA on file R7_753
20:11 Beam stop due to SRC issue. BRIKEN/BigRIPS runs stopped
20:17 Beam is back (49 pnA)
BRIKEN starts run 097
BigRIPS starts run 3103
AIDA on file R7_769
21:00 There have been frequent interruptions of beam (FRC/SRC issues) during the last two to three hours.
Usually resolved after a few minutes, with the beam continuing with good current.
21:12 BRIKEN starts run 098
BigRIPS starts run 3104
AIDA on file R7_782
Screenshot for stats and temperature during R7_783 attached (att. 2 and 3)
Screenshot for energy spectra of FEE12 and 9 as attachment 4 and 5 (around run R7_785)
21:58 BRIKEN starts run 099
BigRIPS starts run 3105
AIDA on file R7_793
23:12 BRIKEN starts run 100
BigRIPS starts run 3106
AIDA on file R7_810 |
Wed Feb 17 11:54:50 2016, CG, TD, 17 February 2016 6x
|
File R17 -> pulser walkthrough.
- Started ~10.00
- Pulser peak amplitude taken from 9.0V to 1.0 V in 1.0V steps
- Shaping time 8us, slow disc threshold 0x8
- File stopped 10.32. Spectra (.msf) saved 10.34
File R18 -> pulser walkthrough.
- Started 10.41.
- Stepped the pulser amp by 0.1V - too small.
- Not useful
- Stopped 10.50
File R19 - pulser walkthrough.
- Started 10.51
- empty
File R20 - empty
File R21 -> pulser Walkthrough
- started 11.18
- 1.0 - 0.2V in 0.2V steps.
- Stopped 11.50. Saved histograms 11.51.
File R22 -> source spectrum.
- slow comp threshold @ 0x7
- forgot to turn off pulser. => no good.
file R23 -> source spectrum.
- slow comp threshold @ 0x7
- started 16.09
18.05 DAQ found stalled, nnaida11 non-responsive
FORCE reset nnaida11 OK
but SETUP and FORCE setup failed with timeouts
Power-cycled all FEE modules to restart
eventual restart following workstation restart
file R29
start 18.50 stop: 09.55 18.2.16
spectra saved to disk 10.55 BST (sic)
see attachments 1-6
09.56 bias +200V I_L +6.105uA ambient temperature 20.1 deg C |
Tue Feb 16 17:32:21 2016, CG, TD, 16 February 2016
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.150uA, ambient temperature +20.1 deg C
DSSSD - AIDA adaptor PCB cabling
4 off LH Coupler (Kapton PCB, 5cm), 2x 34-way Samtec ribbon cable (45cm), RH coupler(Kapton PCB, 10cm)
+ 3M 1245 1.4mil copper foil screen ribbon cables + RH coupler only (i.e. not LH coupler)
+ drain wires -> gold-plated Lemo-00 test input connectors
nnaida 11 & 12 AIDA adaptor PCB rev B
nnaida 13 & 14 AIDA adaptor PCB rev C (LK1 fitted)
ground links LK3 & LK7 fitted to nnaida11-14 AIDA adaptor PCBs
Heavy duty copper cable connects copper front end frames of FEE modules
Nitto 5011N conductive gasket between FEE module and front end frames
Standard ASIC settings
nnaida 11 & 12 - negative input
nnaida 13 & 14 - positive input
Pulser BNC PB-5
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 5.00000V
Polarity +
Pulse top Tail
Atten 10x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier
207Bi source, approx centred on DSSSD, approx 3cm from DSSSD
File R16
shaping time 8uS
slow comparator 8 (dec)
start: 16.2.16 13.15 stop: 09.30 17.2.16
spectra saved to disk 10.30 BST (sic)
Pulser OFF
16.2.16
09.35 Bias +200V, I_L +=5.940uA, ambient temperature +20.0 deg C |
Sun Jul 16 09:36:40 2023, TD, 15-16 July 2023 40x
|
Saturday 15 July
Cooling water ON - temperature c. +19 deg C ( controls adjacent to S4 entrance )
Flow taps ON ( S4 )
Cannot see float in mechanical flow gauge - sediment?
USB-controlled ac mains relay flow interlock ON.
Flow vibration can be felt in cooling manifold and lines.
13x FEE64s installed - see attachment 2 ( courtesy NH )
FEE64 # Cable set #
1 1
2 2
3 3
4 4
5
6 6
7
8 8
9 9
10 5
11 11
12 7
15
10x FEE64s configured - update /MIDAS/config/TclHttpd/aidas-gsi@8015/startup.tcl - see attachment 1
ttyUSB1 aida01
ttyUSB3 aida12
ttyUSB4 aida02
ttyUSB6 aida03
ttyUSB7 aida04
ttyUSB9 aida11
ttyUSB10 aida09
ttyUSB12 aida10
ttyUSB13 aida08
ttyUSB15 aida06
Update /etc/dhcp/dhcpd.conf - correct MAC address aida09
from
host aida09 {
hardware ethernet d8:80:39:41:ee:01;
fixed-address 192.168.11.9;
option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida09";
}
to
host aida09 {
hardware ethernet d8:80:39:41:ee:10;
fixed-address 192.168.11.9;
option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida09";
}
5x MACBs installed - see attachment 3
Note configuration setting 0x0 for all 5x MACBs ( cf. 0x3 per https://elog.ph.ed.ac.uk/DESPEC/517 )
17.30 System wide checks - see attachments 4-7
OK *except* aida09 clock and WR decoder errors
FEE64 temperatures OK - see attachment 8
Sunday 16 July
11.30 Restart
FEE64 temperatures OK - attachment 9
System wide checks - attachments 10-15
aida09 clock errors - no WR timestamp
aida09 WR status 0xd - attachments 16-18
Local controls aida01 cf. aida09 - attachments 19-20
status register 0xf cf. 0x4
BuTIS interface control 0x3 cf. 0x0
14.00 ASIC settings 2021Apr29-13-16-00
slow comparator 0x64
FEE64s 1, 3, 9, 10, 11, 12 + input polarity
FEE64s 2, 4, 6, 8 - input polarity
BNC PB-5 pulser
amplitude 1.0V
attenuation x1
tail pulse
tau_d 1ms
frequency 100Hz
polarity +
EG&G Ortec 534 Sum & Invert Amplifier
to provide - polarity test inputs
Test + and - signals daisy-chained to FEE64s - end of daisy-chain terminated by 50 Ohms.
Adaptor PCB
AIDA Half Rev B 180623 fitted FEE64s 1, 2, 3, 4, 9, 10, 11, 12
*all* LKs fitted
AIDA Half Rev A 120418 fitted FEE64s 6, 8
*no* LKs fitted
No drain wire connections from adaptor PCB to FEE64 cooling plate fitted
per FEE64 rate spectra - attachment 21
p+n FEE64 1.8.L spectra - attachments 22-23
aida01 1.8.L pulser peak width 15 ch FWHM
n+n FEE64 1.8.L spectra - attachments 24-25
double peak spectra
p+n FEE64 1.8.W spectra 20us FSR - attachments 26-28
n+n FEE64 1.8.W spectra 20us FSR - attachments 29-31
Observe significantly higher noise for rev A adaptors
No data observed for aida09
aida01 2.5.* data missing
FEE64 statistics aida01, aida09, good events, ADC data, pause, resume, disc, correlation & WR 48-63 - attachments 32-40 |
Fri Nov 13 16:17:14 2015, CG, TD, 13 Nov 2015 - new Kapton cable tests at DL 6x
|
Powered system up with 4 bare 45cm kaptons connected and nothing seems to have changed from where we left last night
Waveforms in nnaida11 look good and peak widths in nnaida11 1.8.L and 1.10.L are ~70ch.
Turned off flash ADCs for all FEEs by setting Q8_transfer.vhd: 0xa00, offset 0x00, to 0 to assess any affect they may play in the observed noise
no obvious change in peak widths for nnaida11
turned all back on by resetting value of offset 0x00 to 1
Measured RMS deviation of waveform baseline for nnaida11 with 45cm unshielded kapton connected by sampling waveform every 10 samples over 100 samples, i.e. every 200ns for 2us.
1.4.W - mean: 8612.9 st. dev.: 20.5
1.6.W - mean: 8548.9 st. dev.: 16.6
1.8.W - mean: 8900.8 st. dev.: 16.2
1.10.W - mean: 8744.5 st. dev.: 15.0
Removed kapton to check intrinsic RMS deviation of pre-amp using the same method
1.4.W - mean: 8666.8 st. dev.: 4.7
1.6.W - mean: 8586.6 st. dev.: 2.0
1.8.W - mean: 8918.7 st. dev.: 3.2
1.10.W - mean: 8782.3 st. dev.: 4.2
These results are consistent with the pulser peak widths observed with/without Kapton PCBs attached
Make a makeshift screening plate for the adapter PCBs using the Cu tape -> widths in nnaida11 all slightly worse but still comparable.
Connected detector using 4 45cm kaptons (gnd via nnaida13) and baised via nnaida12+13 to +200V with a leakage current of 5.16uA (Room temp 24.2oC)
Initially observed very noisy waveforms but grounding adapter PCBs together improved the noise slightly but not to an acceptable level (see attachment 2+3).
Measure the peak width as a function of shaping time:
t_shape: 0.5 | 1.0 | 2.0 | 4.0 | 8.0
FWHM (ch): 308 | 352 | 345 | 252 | 130
Throughout, the peaks were non-Gaussian and double peaked with single peak widths ~50-70ch.
Pickup effect rather than white noise?
Shielded kapton to nnaida11 with Cu tape and found noise to decrease slightly when grounded to LEMO connector on adapter PCB - bursts reduced in amplitude but persistent periodic noise remained (see attachments 4+5).
Changed grounding around (adapters to NIM crate, adapters to detector box etc) and observed no noticeable difference.
However, removing jumpers grounding PCBs to FEE cards and the grounding between the adapter PCBs makes the noise worse.
Shielded 3 more 45cm kaptons and connected detector -> no leakage current.
Found several pins on 2 kaptons to be broken. Will get these fixed in Edinburgh for use next week.
Reconnected old kaptons and system is as expected - slower rise time in waveforms but reasonable levels of noise and expected peak widths (see attachment 6). |
Sat Nov 14 20:19:43 2015, CG, TD, PJCS, 13 Nov 2015 - new Kapton cable tests at DL
|
| Quote: |
|
Powered system up with 4 bare 45cm kaptons connected and nothing seems to have changed from where we left last night
Waveforms in nnaida11 look good and peak widths in nnaida11 1.8.L and 1.10.L are ~70ch.
Turned off flash ADCs for all FEEs by setting Q8_transfer.vhd: 0xa00, offset 0x00, to 0 to assess any affect they may play in the observed noise
no obvious change in peak widths for nnaida11
turned all back on by resetting value of offset 0x00 to 1
Measured RMS deviation of waveform baseline for nnaida11 with 45cm unshielded kapton connected by sampling waveform every 10 samples over 100 samples, i.e. every 200ns for 2us.
1.4.W - mean: 8612.9 st. dev.: 20.5
1.6.W - mean: 8548.9 st. dev.: 16.6
1.8.W - mean: 8900.8 st. dev.: 16.2
1.10.W - mean: 8744.5 st. dev.: 15.0
Removed kapton to check intrinsic RMS deviation of pre-amp using the same method
1.4.W - mean: 8666.8 st. dev.: 4.7
1.6.W - mean: 8586.6 st. dev.: 2.0
1.8.W - mean: 8918.7 st. dev.: 3.2
1.10.W - mean: 8782.3 st. dev.: 4.2
These results are consistent with the pulser peak widths observed with/without Kapton PCBs attached
Make a makeshift screening plate for the adapter PCBs using the Cu tape -> widths in nnaida11 all slightly worse but still comparable.
Connected detector using 4 45cm kaptons (gnd via nnaida13) and baised via nnaida12+13 to +200V with a leakage current of 5.16uA (Room temp 24.2oC)
Initially observed very noisy waveforms but grounding adapter PCBs together improved the noise slightly but not to an acceptable level (see attachment 2+3).
Measure the peak width as a function of shaping time:
t_shape: 0.5 | 1.0 | 2.0 | 4.0 | 8.0
FWHM (ch): 308 | 352 | 345 | 252 | 130
Throughout, the peaks were non-Gaussian and double peaked with single peak widths ~50-70ch.
Pickup effect rather than white noise?
Shielded kapton to nnaida11 with Cu tape and found noise to decrease slightly when grounded to LEMO connector on adapter PCB - bursts reduced in amplitude but persistent periodic noise remained (see attachments 4+5).
Changed grounding around (adapters to NIM crate, adapters to detector box etc) and observed no noticeable difference.
However, removing jumpers grounding PCBs to FEE cards and the grounding between the adapter PCBs makes the noise worse.
Shielded 3 more 45cm kaptons and connected detector -> no leakage current.
Found several pins on 2 kaptons to be broken. Will get these fixed in Edinburgh for use next week.
Reconnected old kaptons and system is as expected - slower rise time in waveforms but reasonable levels of noise and expected peak widths (see attachment 6).
|
To power down the ADCs the register to be loaded with 0xFF is in the "Local Registers" window.
What you did was to disable the readout of the ADCs which doesn't change the level of activity or reduce the power.
|
Thu Nov 12 18:28:41 2015, CG, TD, 12 Nov 2015 - new Kapton cable tests at DL 23x
|
This morning PC-S changed firmware and DAQ software to that of the RIKEN setup
Firmware version /MIDAS/AIDA/FEE_Riken_Apr14_26.bin
DAQ software has been changed from AidaExecV8 to AidaExecV7
nnaida12 is not playing ball -> ASIC4 has stopped outputting any data and ASICs2+3 are missing multiple channels. Not running hot
Suspicions over the power cable as the source of the problem (previous nnaida12 developed noisy ASICs) and tests found this not to be an issue. Neither are the MACB HDMIs as they were swapped for new ones last night.
Then changed the adapted PCB to see if that was a possible cause. No change -> FEE or mezzanine problem?
With old kaptons attached, peak widths in nnaida11 are ~180-190ch with shaping time 2us. Increasing this to 6us brings FWHM to ~170ch, and increasing further to 8us reduces the FWHM to ~140ch.
FAO PC-S -> After lunch, upon clearing the spectra, nnaida12 stopped responding (terminal screenshot below - attachment 19). Brought system to halt. Power cycling resolved the isssue.
A series of tests over the afternoon found peak widths to vary only by ~10-20% when the second length (detector end) of kapton was removed. Removing the connecter PCB showed no improvement over this.
However, removing the Cu tape put on as shielding, reduces the FWHM by a factor of 3.
see nnaida11_wav_newCable.png (attachment 17) for before waveforms -> the Cu tape is clearly doing something we are unaware of.
Attachment 18 shows improvement in just removing the Cu tape.
Re-adding connector PCB, increased widths by ~5-10% and re-adding second length of kapton increased widths by x3-4.
Second length of kapton was tested on its own and was found to have roughly the same widths as the first length, no issues with this cable -> increase in FWHM just due to increased capacitance (possibly...)
Connected a 45cm kapton to each FEE, 3 with the ground connection broken, and the grounded one connected to nnaida13 (attachment 23).
Varied shaping time and found 0.5us to consistently provide lowest FWHM (summary graphs to follow).
With just one length (+connector PCB) of kapton connected to adapter PCBs, waveforms all look very good, too good to explain the higher than expected FWHM.
Upon connecting 90cm of kaptons, noise creeps in to waveforms. Also observed reoccurring periodic transients in some waveforms (attachment 20).
Rise time with new kaptons connected (45cm) noticeably faster that with the old cables -> new show rise time of ~140ns, old cables has rise time ~3us.
To do tomorrow:
connect detector to FEEs using 4 45cm kaptons, assess affect on FWHM. |
Wed Nov 11 17:36:19 2015, CG, TD, 11 Nov 2015 - new Kapton cable tests at DL 18x
|
Upon starting system with nothing connected beyond adapter PCBs, observed pulser peak FWHM of ~12-15ch in nnaida13+14 and ~15-18ch in nnaida11+12
All newly shielded new kapton cables passed continuity tests between drain wire and Cu tape.
Throughout all tests, grounded cable has been connected to nnaida13. All other nnaida have ground isolated cables.
Initially, upon biasing, leakage current reached 50uA at ~2V HV and tripped.
After much searching, found one of the drain wires from the kaptons to be touching the core of the HV cable.
Tidied all drain wires and made sure none were touching anything they shouldn't.
Bias ramped up successfully to +200V with leakage current of 5.630uA @ room temp ~25.5oC.
Very noisy waveforms to begin with, notes Cu tape on kaptons was in contact with the Al housing for the detector.
Isolated kaptons from Al housing with black tape and the waveforms became much less noisy, but still not as calm as expected.
Connecting Cu covered kaptons with croc clips made this problem worse. However, grounding adapter PCBs together improved the waveform noise.
FWHM with new cables ~200-300ch, much worse than previous cables.
Reconnected system with old kaptons and they now see FWHM of ~200-300ch, as opposed to ~100ch observed earlier in the year.
What has changed?
Waveforms are smoother with old kaptons, but still not as flat as previously observed.
Had some issues with waveforms not being output by FEE cards and ASIC4 of nnaida12 has failed to Check/Load and is not outputting any data.
Standard ASIC parameters throughout, with shaping time of 0.5us. |
Sat Jun 10 15:56:31 2017, ZLiu, 0:00-8:00 11 June 2017 46x
|
00:35 Stop DAQ
AIDA file RIBF128/R7_1172
00:36 Start DAQ
AIDA file RIBF128/R7_1172
BRIKEN file 118
BigRIPS file 4008
00:50 Detector biases & leakage currents OK - see attachments 4, 5
FEE64 Temperatures OK - see attachments 3
good events statistics stable & OK - see attachment 2
No warnings/errors reported by merger since DAQ GO - see attachment 1
*all* system wide checks OK
01:33 Stop DAQ
AIDA file RIBF128/R7_1187
01:36 Start DAQ
AIDA file RIBF128/R7_11 87
BRIKEN file 119
BigRIPS file 4009
02:32 Stop DAQ
AIDA file RIBF128/R7_1202
02:34 Start DAQ
AIDA file RIBF128/R7_1202
BRIKEN file 120
BigRIPS file 4010
2:40 Detector biases & leakage currents OK - see attachments 6, 7
FEE64 Temperatures OK - see attachments 8
good events statistics stable & OK - see attachment 9
No warnings/errors reported by merger since DAQ GO - see attachment 10
implantation profile - see attachment 11
*all* system wide checks OK
03:33 Stop DAQ
AIDA file RIBF128/R7_1217
03:34 Start DAQ
AIDA file RIBF128/R7_1217
BRIKEN file 121
BigRIPS file 4011
03:40 Detector biases & leakage currents OK - see attachments 12, 13
FEE64 Temperatures OK - see attachments 14
good events statistics stable & OK - see attachment 15
No warnings/errors reported by merger since DAQ GO - see attachment 16
implantation profile - see attachment 17
*all* system wide checks OK
04:33 Stop DAQ
AIDA file RIBF128/R7_1232
04:34 Start DAQ
AIDA file RIBF128/R7_1232
BRIKEN file 122
BigRIPS file 4012
04:40 Detector biases & leakage currents OK - see attachments 18, 19
FEE64 Temperatures OK - see attachments 20
good events statistics stable & OK - see attachment 21
No warnings/errors reported by merger since DAQ GO - see attachment 22
implantation profile - see attachment 23
DSSD rates - see attachment 24
*all* system wide checks OK
05:33 Stop DAQ
AIDA file RIBF128/R7_1248
05:34 Start DAQ
AIDA file RIBF128/R7_1248
BRIKEN file 123
BigRIPS file 4013
05:40 Detector biases & leakage currents OK - see attachments 25, 26
FEE64 Temperatures OK - see attachments 27
good events statistics stable & OK - see attachment 28
No warnings/errors reported by merger since DAQ GO - see attachment 29
implantation profile - see attachment 31
DSSD rates - see attachment 30
*all* system wide checks OK
05:52 beam stopped due to RF problem of the cyclotron
05:58 beam back
06:30 Stop DAQ
AIDA file RIBF128/R7_1262
06:31 Start DAQ
AIDA file RIBF128/R7_1262
BRIKEN file 124
BigRIPS file 4014
06:50 Detector biases & leakage currents OK - see attachments 32, 33
FEE64 Temperatures OK - see attachments 34
good events statistics stable & OK - see attachment 35
No warnings/errors reported by merger since DAQ GO - see attachment 36
implantation profile - see attachment 37
DSSD rates - see attachment 38
*all* system wide checks OK
07:30 Stop DAQ
AIDA file RIBF128/R7_1278
07:32 Start DAQ
AIDA file RIBF128/R7_1278
BRIKEN file 125
BigRIPS file 4015
07:50 Detector biases & leakage currents OK - see attachments 39, 40
FEE64 Temperatures OK - see attachments 41
good events statistics stable & OK - see attachment 42
No warnings/errors reported by merger since DAQ GO - see attachment 43
implantation profile - see attachment 44
DSSD rates - see attachment 45
??? *all* system wide checks: problem with AIDA17 ? -- see attachment 46
*all* system wide checks again, looks OK now
|
Wed Oct 1 18:37:42 2014, Chris Griffin, 01/10/2014 Tests @ DL 11x
|
Kept largely the same settings at the entry https://elog.ph.ed.ac.uk/AIDA/4
Same ASIC settings
Emco used to provide HV
HV+ --> nnaida11 & nnaida12 (p+n strips, -ve pulser input)
HV- --> nnaida13 & nnaida14 (n+n strips, +ve pulser input)
Stat spectra look roughly the same as those included in the above entry.
Now see large peak around ~32k, present in all spectra to some degree. Mostly larger than corresponding pulser peak.
Power cycled several times, reloaded+checked ASIC settings etc., no change. Any ideas? Have I forgotten something?
The ASIC events stats rate is also very high on nnaida11,12,14. 13 is high but not excessively.
Pulser peak shows double peaking on all ASIC channels with individual widths of ~250ch and total width of ~500ch.
Noise peak has width ~50ch. This has reduced somewhat (was ~100s ch) with my last power cycle, but clearly still not gone.
I'll try again in the morning, but if no change will proceed with the shaping time tests with the spectra as they are.
|
Mon Jun 6 13:46:45 2016, TD, ML & YS, /var/log/messages
|
aidas1 log |
Wed Nov 22 15:12:40 2023, TD, /tmp/log8015_error
|
| |
Tue Apr 28 05:59:48 2015, AE, "beta" trigger from MACB Trigger outputs 6x
|
We've constructed a beta trigger with the signals from the p-side of the three DSSDs (OR of all fast
discriminator channels of each FEE).
det1 -> NNAIDA#6 -> MACB#10.2
det1 -> NNAIDA#10 -> MACB#3.4
det2 -> NNAIDA#2 -> MACB#10.3
det2 -> NNAIDA#14 -> MACB#3.2
det3 -> NNAIDA#1 -> MACB#10.4
det3 -> NNAIDA#13 -> MACB#3.3
beta-trigger is OR of all above MACB trigger outputs. Diagram for logic circuit is in paper logbook.
Bellow are some oscilloscope sceenshots. Some things to note about the MACB trigger outputs:
- all show multiple outputs for each pulser input. Some channels seem to have higher level for such 'multiplicity'.
- the width of the outputs varies for different channels (from ~40 ns to ~300 ns)
- the delay between the pulser signal and the output from the MACB is quite large, and also different for each
MACB (measured from ~800 ns up to 2 usec).
Scope screenshots:
- light blue is the trigger output from Pulser
- dark blue is output from trigger channel of MACB
MACB#10.2: tek00001.tif
MACB#3.4: tek00002.tif
MACB#10.3: tek00005.tif
MACB#3.2: tek00007.tif
MACB#10.4: tek000011.tif
MACB#3.3: tek000018.tif |
|