Tue Nov 8 11:51:28 2016
Author
*
:
Subject
*
:
Some spectra show spikes or do not look normal. This should be checked. Look at BRIKEN, Ring4, Ring5, Ring6, Ring7, D4, G7, Clovers, Rate1, Rate2, Rate3, etc. Reset time stamp: 21.00, 8th-nov. Probably there was a problem with the clock from AIDA related to the LN2 filling. So to recover, we need to stop acq (only stop acquisition in DACQ control), send to hard (send parameters again) and once all the parameters were sent, Send Sync pulse.