> Check of calibrations and rates for neutron detectors with Conf File 161107Conf_BrikenFull.xlsx > > Measurement 161122_2208_003.root has been used to check the calibrations and counting rates in the BRIKEN > detector. Condition of the measurement: > 252Cf at 22.5cm from moderator, opposite sede from AIDA > > RESULTS > > Calibrations factors: > Calibrations of the neutron detectors are reasonable in the current conditions. A check detector by detector > (see attached file) shows that the full peak for most of the counters is around 764keV. Only 3 counters are > shifted more than 10keV: > He134 741.901 > He135 750.312 > He140 748.587 > > The impact on the counting rates would not be serious. > > Counting rates (n/pulser): > ######################################### > Total neutron rate: 1516.28 > ######################################### > BY RING: > ######################################### > R1 359.057 > R2 222.997 > R3 386.299 > R4 253.833 > R5 143.434 > R6 90.6357 > R7 60.025 > > The total current counting rate is lower than rate obtained in Oct 12th (total 1550.91 n/pulser). The difference > is due mainly to a decrease in the counting rates for rings 1 to 4.