<p>After experiment 133Ba calibration</p>
<p>Working directory: /data/1706FinalCalibrations<br />
Configuration file: 170604_Conf_BrikenFull_AfterCrisis.xlsx</p>
<p>DLT file name: 170613_1601_133Ba_002.dlt<br />
Start: 16:02<br />
Stop: 12:25 (2017/06/14)<br />
Online rootfile: 133Ba_cal_z0_1602_1227nextday_170613.root</p>
<p>COMMENTS:</p>
<p>1. Source centered in both BRIKEN and clovers (faced towards clover D4)<br />
2. No AIDA in<br />
3. D4 clover pulsers are at 253, 260, 265, and 271 keV<br />
4. G7 clover pulsers are beyond 5 MeV. No pulser for crystal G7Green</p>
<p>5. During the lasts dlt files (170613_1601_133Ba_009.dlt and 170613_1601_133Ba_010.dlt) the source was removed from the center of BRIKEN.</p>
<p> </p> |