AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  BRIKEN  ELOG logo
Message ID: 341     Entry time: Sun Oct 22 08:52:17 2017
Author: KR,JA,ATD,JLT 
Subject: Clover resolution 
Resolution check using 40K 1460 keV line 
in the background measurement overnight
(File: 171022_0126_1101_bck.root)

D4: 2.9 1keV @ 1461.8 keV
G7: 2.98 keV @ 1460.0 keV
ELOG V3.1.4-unknown