Message ID: 396     Entry time: Fri Oct 26 03:31:33 2018
Author: AT2 & JA 
Subject: Test pulser effect on baseline 
We have tested possible change of the baseline in the digitizer due to the pulser connection. NO EFFECT has been
found!


Test for detector He086 ORNL:
He086_NoPulser_samples.png  -> Without pulser, baseline around 1000
He086_Pulser_samples.png    -> With pulser, baseline around 1000


Test for detector He001 UPC:
He001_Pulser_samples.png  -> With pulser, baseline around 1200
He001_NoPulser_samples.png  -> Without pulser, baseline around 1200


Test for detector He013 RIKEN:
He013_Pulser_samples.png  -> With pulser, baseline around 850
He013_NoPulser_samples.png  -> Without pulser, baseline around 850
Attachment 1: He086_NoPulser_samples.png  20 kB  Uploaded Fri Oct 26 04:36:33 2018  | Hide | Hide all
He086_NoPulser_samples.png
Attachment 2: He086_Pulser_samples.png  19 kB  Uploaded Fri Oct 26 04:39:25 2018  | Hide | Hide all
He086_Pulser_samples.png
Attachment 3: He001_Pulser_samples.png  18 kB  Uploaded Fri Oct 26 04:43:35 2018  | Hide | Hide all
He001_Pulser_samples.png
Attachment 4: He001_NoPulser_samples.png  19 kB  Uploaded Fri Oct 26 04:45:44 2018  | Hide | Hide all
He001_NoPulser_samples.png
Attachment 5: He013_Pulser_samples.png  20 kB  Uploaded Fri Oct 26 04:51:06 2018  | Hide | Hide all
He013_Pulser_samples.png
Attachment 6: He013_NoPulser_samples.png  21 kB  Uploaded Fri Oct 26 04:51:18 2018  | Hide | Hide all
He013_NoPulser_samples.png