We have tested possible change of the baseline in the digitizer due to the pulser connection. NO EFFECT has been
found!
Test for detector He086 ORNL:
He086_NoPulser_samples.png -> Without pulser, baseline around 1000
He086_Pulser_samples.png -> With pulser, baseline around 1000
Test for detector He001 UPC:
He001_Pulser_samples.png -> With pulser, baseline around 1200
He001_NoPulser_samples.png -> Without pulser, baseline around 1200
Test for detector He013 RIKEN:
He013_Pulser_samples.png -> With pulser, baseline around 850
He013_NoPulser_samples.png -> Without pulser, baseline around 850 |