AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  BRIKEN  ELOG logo
Message ID: 401     Entry time: Tue Oct 30 01:01:42 2018
Author: JLT, AA 
Subject: Adjustment of 3He thresholds 
Background with BigRIPS+ZD already working
No pulser

After adjustment of thresholds the rate of BRIKEN noise
was reduced to 4.5kHz
The corresponding configuration file is:
181030_Conf_BrikenFull_LowThrAdj_masterCLK.xlsx
To be decided if we want to reduce further the noise rate

We also changed the calibration of the F11 plastic to have it in the histogram range
F11_PL_R_LG: 2.73 -> 1.0 
F11_PL_L_LG: 2.32 -> 2.0


Further adjustment of few channels: Now noise rate is 2.7kHz
(2018/10/30 15:45)
Attachment 1: 181030_Conf_BrikenFull_LowThrAdj_masterCLK.xlsx  83 kB  Uploaded Tue Oct 30 07:17:35 2018
Attachment 2: Screenshot_from_2018-10-30_16-22-32.png  33 kB  Uploaded Tue Oct 30 07:22:46 2018  | Hide | Hide all
Screenshot_from_2018-10-30_16-22-32.png
ELOG V3.1.4-unknown