|
AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
BRIKEN |
|
|
Message ID: 406
Entry time: Mon Nov 5 06:46:30 2018
|
Author: |
KR, YR, MW, JLT, ... |
Subject: |
Clover check |
|
|
With detectors still on the floor we check signals and found
that D4 gives no signal with PA power. Two power supplies/cables were checked.
The detector was still working just before Jungclaus experiment. In July it was working.
Temperature monitors OK: 76-80 units
We mount both CLovers in position and make all connections.
We apply HV and check the resolution of G7.
Source: 152Eu
File: 181105_1550_CloverCheck_152Eu.root
As can be seen in the attached figure the calibration used last year
is still pretty good
Resolution @ 1408 keV
G7 black: 2.75keV
NO. NAME VALUE ERROR SIZE DERIVATIVE
1 Constant 6.57807e+02 1.35347e+01 4.63240e-02 -7.68897e-07
2 Mean 1.40766e+03 1.96681e-02 6.71225e-04 5.14533e-03
3 Sigma 1.16983e+00 1.52231e-02 1.41407e-05 -1.28285e-02
G7 red: 2.97keV
NO. NAME VALUE ERROR SIZE DERIVATIVE
1 Constant 5.60626e+02 1.17448e+01 3.69357e-02 1.06287e-06
2 Mean 1.40659e+03 2.29862e-02 6.70713e-04 1.36869e-04
3 Sigma 1.26305e+00 1.62145e-02 1.30239e-05 -1.44151e-02
G7 green: 2.75keV
NO. NAME VALUE ERROR SIZE DERIVATIVE
1 Constant 7.87830e+02 1.48889e+01 3.68407e-02 5.57077e-06
2 Mean 1.40597e+03 1.75988e-02 6.70417e-04 -2.47334e-03
3 Sigma 1.17167e+00 1.41220e-02 9.59214e-06 1.55873e-02
G7 blue: 2.73keV
NO. NAME VALUE ERROR SIZE DERIVATIVE
1 Constant 4.99089e+02 1.19364e+01 -6.77351e-03 -4.15987e-06
2 Mean 1.40489e+03 2.17210e-02 2.57134e-05 2.38923e-03
3 Sigma 1.16057e+00 1.73778e-02 3.14777e-06 2.01018e-02
FWHM: |
|
|
|