AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
BRIKEN
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Mon Mar 31 13:09:00 2025
Author
*
:
Subject
*
:
> Adjustment of HV and PA gain > > We adjust the HV for each tube type and the gain of the preamplifier (PA) in order > to have a similar signal amplitude for the 764keV peak in all of them. > Because of the very different gain between tubes we are not able to combine in the same > PA tubes from RIKEN and UPC, thus we have to go for 10 PA instead of 9: > > Settings: Preamp (type), Tube type, HV, Gain, height or 764keV pulse, pulse polarity > PA1 (D), RIKEN, 1450V, LowGain, ~400mV, POS > PA2 (D), RIKEN, 1450V, LowGain, ~400mV, POS > PA3 (D), UPC, 1550V, HighGain, ~400mV, POS > PA4 (D), UPC, 1550V, HighGain, ~400mV, POS > PA5 (U), UPC, 1550V, HighGain, ~600mV, NEG > PA6 (U), ORNL1, 1200V, LowGain, ~400mV, NEG > PA7 (U), ORNL2, 1750V, LowGain, ~400mV, NEG > PA8 (U), ORNL2, 1750V, LowGain, ~400mV, NEG > PA9 (U), ORNL2, 1750V, LowGain, ~400mV, NEG > PA10 (U), ORNL2, 1750V, LowGain, ~400mV, NEG > > We need to use two different pulse generators to adjust differently the pulse height > of UPC tubes