<p>We observed 6 pm LN2 filling of both clovers on Monday Nov 5th.</p>
<p>D4 went full within reasonable time of below 10 minutes.</p>
<p>Filling of G7 was stopped by "emergency sensor" after about 15 minutes of filling.</p>
<p>Likely the temp sensor at the LN2 overflow pipe of G7 should be changed.</p>
<p>The clock at LN2 fill system is off by about 25 minutes (filling comes later than set at the control unit.)</p>
<p> </p>
</td>
</tr>
</tbody>
</table>
<p>"The clock at LN2 fill system is off by about 25 minutes (filling comes later than set at the control unit.)"</p>
<p>Yes, This is how it always was.</p>
<p> </p>
<p>"Filling of G7 was stopped by "emergency sensor" after about 15 minutes of filling. Likely the temp sensor at the LN2 overflow pipe of G7 should be changed."</p>
<p>You can put this on a different line of the fill system....</p> |