Entry time:
Wed Jul 20 16:40:57 2016
Author
*
:
Subject
*
:
Calibration and resolution check CLOVER detectors were calibrated in energy with 60Co and 137Cs 60Co source: 2530, 8.81E4Bq (16/03/2010) 137Cs source: CD393 (box), ? Sources were placed at the center Resolution(FWHM@1333keV): G7 Blue:2.67keV G7 Green: 2.84keV G7 Red: 2.77keV G7 Black: 2.64keV D4 Blue: 3.60keV D4 Green: 3.35keV D4 Red: 3.41keV D4 Black: 4.85keV (!?) Efficiency calibration with 152Eu Source place at the center Source: 272 (09-7011), 1.062E4Bq (error:1.9%)@16/03/2010 File: 160719_2201_152Eu_028.dlt Online: 160719_2201_152Eu_028.root Attached 152Eu spectrum
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
152Eu.png